CLASS II ONGOING Device recall · Reported 29 Jul 2020 · FDA Enforcement Report

Computed Tomography X-ray System recalled by Shanghai United Imaging

Shanghai United Imaging Healthcare is recalling Computed Tomography X-ray System, distributed in Texas. The recall was initiated on 23 May 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Numbers: 353020, 353021, and 353023 with software version R001.11.0.729741-Re-1100
Quantity recalled3

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2641-2020
FDA event ID
85771
Recalling firm
Shanghai United Imaging Healthcare, Shanghai, N/A
Classification
Class II
Status
Ongoing
Recall initiated
23 May 2020
FDA classified
17 Jul 2020
Reported
29 Jul 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

23 May 2020Recall initiated by company
17 Jul 2020Classified by FDA+55 days
29 Jul 2020Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.

Product as listed by the FDA

Computed Tomography X-ray System, Model: uCT 550, GTIN: 06971576831036 Intended use The uCT Computed Tomography X-ray System uCT 530/uCT 550 is a computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck and vascular).

Where it was distributed

US Distribution to state of: TX

Texas

Questions about this recall

Why was it recalled?

Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.

Which lots are affected?

Serial Numbers: 353020, 353021, and 353023 with software version R001.11.0.729741-Re-1100

Where was it sold?

US Distribution to state of: TX

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Shanghai United Imaging Healthcare

All 33

Other Heart & Vascular Devices recalls

All 6,217