CLASS II Device recall · Reported 8 Feb 2017 · FDA Enforcement Report

Cryptococcal Antigen Lateral Flow Assay (CrAg LFA) recalled

Immuno-Mycologics is recalling Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003, distributed nationwide in the US. The recall was initiated on 27 Dec 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes161028133653JH, 161003111544KH, and 161007105036JG
Quantity recalled630 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1109-2017
FDA event ID
76159
Recalling firm
Immuno-Mycologics, Norman, OK
Classification
Class II
Status
Terminated
Recall initiated
27 Dec 2016
FDA classified
27 Jan 2017
Reported
8 Feb 2017
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Potency & Assay

Timeline

27 Dec 2016Recall initiated by company
27 Jan 2017Classified by FDA+31 days
8 Feb 2017Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003

Where it was distributed

Nationwide throughout the US

Nationwide

Questions about this recall

Why was it recalled?

The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.

Which lots are affected?

161028133653JH, 161003111544KH, and 161007105036JG

Where was it sold?

Nationwide throughout the US

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.