CLASS II Device recall · Reported 18 May 2016 · FDA Enforcement Report

Dimension Vista Assays: Enzymatic Creatine (ECRE) recalled

Siemens Healthcare Diagnostics is recalling Dimension Vista Assays: Enzymatic Creatine (ECRE), distributed nationwide in the US. The recall was initiated on 17 Mar 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesDevice Listing: D089144 Enzymatic Creatinine (ECRE), Catalog # K1270A
Quantity recalled137422 units total

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1664-2016
FDA event ID
73667
Recalling firm
Siemens Healthcare Diagnostics, Newark, DE
Classification
Class II
Status
Terminated
Recall initiated
17 Mar 2016
FDA classified
11 May 2016
Reported
18 May 2016
Type
Voluntary: Firm initiated
Customers notified by
N/A
Reason category
Potency & Assay

Timeline

17 Mar 2016Recall initiated by company
11 May 2016Classified by FDA+55 days
18 May 2016Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole. NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Dimension Vista Assays: Enzymatic Creatine (ECRE)

Where it was distributed

Nationwide: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico, St. Thomas, VI, Guam, Pacific Mariana Islands

Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMaryland

Questions about this recall

Why was it recalled?

Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole. NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific

Which lots are affected?

Device Listing: D089144 Enzymatic Creatinine (ECRE), Catalog # K1270A

Where was it sold?

Nationwide: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico, St. Thomas, VI, Guam, Pacific Mariana Islands

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

15 other products · FDA event 73667

This product is part of a recall event; the main page for the event is Dimension Assays: Triglycerides (TGL) recalled.

ReportedClassRecallCompanyReasonWhere
CLASS IIDimension Assays: Triglycerides (TGL) recalledSiemens Healthcare DiagnosticsPotency & AssayNationwide
CLASS IIAdvia Chemistry Glucose Oxidase Assay And Advia Chemistry recalledSiemens Healthcare DiagnosticsPotency & AssayNationwide
CLASS IIDimension Assays: Enzymatic Creatinine recalledSiemens Healthcare DiagnosticsPotency & AssayNationwide
CLASS IIDimension Assays: Direct HDL Cholesterol (AHDL) recalledSiemens Healthcare DiagnosticsPotency & AssayNationwide
CLASS IIDimension Vista Assays: Triglycerides (TRIG) recalledSiemens Healthcare DiagnosticsPotency & AssayNationwide
CLASS IIAdvia Chemistry Triglycerides_2 recalledSiemens Healthcare DiagnosticsPotency & AssayNationwide
CLASS IIDimension Vista Assays: Direct HDL Cholesterol (AHDL) recalledSiemens Healthcare DiagnosticsPotency & AssayNationwide
CLASS IIDimension Vista Assays: Uric Acid (URCA) recalledSiemens Healthcare DiagnosticsPotency & AssayNationwide
CLASS IIAdvia Chemistry Assay: Direct HDL Cholesterol (D-hdl) recalledSiemens Healthcare DiagnosticsPotency & AssayNationwide
CLASS IIAdvia Chemistry Assay: Uric Acid and Uric Acid, concentrated recalledSiemens Healthcare DiagnosticsPotency & AssayNationwide
CLASS IIAdvia Chemistry Assays - Cholesterol_2 concentrated recalledSiemens Healthcare DiagnosticsPotency & AssayNationwide
CLASS IIAdvia Chemistry Assay: Lactate recalledSiemens Healthcare DiagnosticsPotency & AssayNationwide
CLASS IIDimension Assays: Uric Acid (URCA) recalledSiemens Healthcare DiagnosticsPotency & AssayNationwide
CLASS IIAdvia Chemistry Assay: LDL Cholesterol, Direct recalledSiemens Healthcare DiagnosticsPotency & AssayNationwide
CLASS IIAdvia Chemistry Enzymatic Creatinine_2 recalledSiemens Healthcare DiagnosticsPotency & AssayNationwide

More recalls from Siemens Healthcare Diagnostics

All 694

Other Test Kits & Diagnostics recalls

All 2,656