CLASS II ONGOING Device recall · Reported 30 Jul 2025 · FDA Enforcement Report

Discovery MR750w 3.0T recalled by GE Medical Systems

GE Medical Systems is recalling Discovery MR750w 3.0T, distributed nationwide in the US. The recall was initiated on 23 May 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesGTIN: 00840682103817 00195278229519
Quantity recalledN/A

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2132-2025
FDA event ID
97030
Recalling firm
GE Medical Systems, Waukesha, WI
Classification
Class II
Status
Ongoing
Recall initiated
23 May 2025
FDA classified
18 Jul 2025
Reported
30 Jul 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

23 May 2025Recall initiated by company
18 Jul 2025Classified by FDA+56 days
30 Jul 2025Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.

Product as listed by the FDA

Discovery MR750w 3.0T

Where it was distributed

Worldwide distribution - US Nationwide and the country of China.

Nationwide

Questions about this recall

Why was it recalled?

GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.

Which lots are affected?

GTIN: 00840682103817 00195278229519

Where was it sold?

Worldwide distribution - US Nationwide and the country of China.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

3 other products · FDA event 97030

This product is part of a recall event; the main page for the event is Signa Architect recalled by GE Medical Systems.

ReportedClassRecallCompanyReasonWhere
CLASS IISigna Architect recalled by GE Medical SystemsGE Medical SystemsOtherNationwide
CLASS IISigna Architect AIR recalled by GE Medical SystemsGE Medical SystemsOtherNationwide
CLASS IISigna PET/MR recalled by GE Medical SystemsGE Medical SystemsOtherNationwide

More recalls from GE Medical Systems

All 160

Other Other Medical Devices recalls

All 13,908