CLASS II Device recall · Reported 10 Jul 2019 · FDA Enforcement Report

Disposable Vitrectomy Cutter recalled by Johnson & Johnson Surgical

Johnson & Johnson Surgical Vision is recalling Disposable Vitrectomy Cutter, distributed in 10 states. The recall was initiated on 24 Sep 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Number: 60084881 Mfg. date: 2017-09-05 Exp. date: 2020-09-05 UDI: (01)05050474534117 (17)200905(10)60084881
Quantity recalled392 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1893-2019
FDA event ID
82868
Recalling firm
Johnson & Johnson Surgical Vision, Santa Ana, CA
Classification
Class II
Status
Terminated
Recall initiated
24 Sep 2018
FDA classified
28 Jun 2019
Reported
10 Jul 2019
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

24 Sep 2018Recall initiated by company
28 Jun 2019Classified by FDA+277 days
10 Jul 2019Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Packaging error, 20GA Vitrectomy Cutter may be found in a 25GA package. Use of a Vitrectomy cutter that is a different size than expected could lead to the need to alter the surgical technique, including corneal incision enlargement.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Disposable 25GA Vitrectomy Cutter, REF NGP0025, RX Only, Abbott Medical Optics Inc., The product provides the item necessary to perform one vitrectomy procedure. It is used to remove vitreous from the eye. It has been designed as a disposable (single use)

Where it was distributed

US Distribution to states of: NC, NV, TX, NY, UT, C, MN, NY, WA, NJ, OH, NV, CA and Internationally to: Australia and Japan.

CaliforniaMinnesotaNorth CarolinaNew JerseyNevadaNew YorkOhioTexasUtahWashington

Questions about this recall

Why was it recalled?

Packaging error, 20GA Vitrectomy Cutter may be found in a 25GA package. Use of a Vitrectomy cutter that is a different size than expected could lead to the need to alter the surgical technique, including corneal incision enlargement.

Which lots are affected?

Lot Number: 60084881 Mfg. date: 2017-09-05 Exp. date: 2020-09-05 UDI: (01)05050474534117 (17)200905(10)60084881

Where was it sold?

US Distribution to states of: NC, NV, TX, NY, UT, C, MN, NY, WA, NJ, OH, NV, CA and Internationally to: Australia and Japan.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Johnson & Johnson Surgical Vision

All 18

Other Other Medical Devices recalls

All 13,857