CLASS II ONGOING Device recall · Reported 27 Aug 2025 · FDA Enforcement Report

Dynex Agility Sample Tips recalled by Dynex Technologies

Dynex Technologies is recalling Dynex Agility Sample Tips, distributed in California, Florida, North Carolina, New Jersey, New York. The recall was initiated on 16 Jul 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Number: 67910; UDI-DI: 5060456180072; Lot Number: 863366;
Quantity recalled237 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2345-2025
FDA event ID
97319
Recalling firm
Dynex Technologies, Chantilly, VA
Classification
Class II
Status
Ongoing
Recall initiated
16 Jul 2025
FDA classified
21 Aug 2025
Reported
27 Aug 2025
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Labeling Errors

Timeline

16 Jul 2025Recall initiated by company
21 Aug 2025Classified by FDA+36 days
27 Aug 2025Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The internal label affixed to the sample tip rack is incorrectly labeled as reagent tip racks. When scanned by the Agility instrument, the tip rack is misidentified as a different tip type, rendering it incompatible with the system and preventing use which causes a delay in obtaining patient results.

Product as listed by the FDA

Dynex Agility Sample Tips; Model Number: 67910; The 67910 Agility Sample Tips are an accessory/consumable used with the Agility Automated ELISA system.

Where it was distributed

Domestic: CA, FL, NC, NJ, NY;

CaliforniaFloridaNorth CarolinaNew JerseyNew York

Questions about this recall

Why was it recalled?

The internal label affixed to the sample tip rack is incorrectly labeled as reagent tip racks. When scanned by the Agility instrument, the tip rack is misidentified as a different tip type, rendering it incompatible with the system and preventing use which causes a delay in obtaining patient results.

Which lots are affected?

Model Number: 67910; UDI-DI: 5060456180072; Lot Number: 863366;

Where was it sold?

Domestic: CA, FL, NC, NJ, NY;

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Dynex Technologies

All 2
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIDynex Agility, Agility Analyzer recalled by Dynex TechnologiesDynex TechnologiesPotency & AssayNationwide

Other Other Medical Devices recalls

All 13,908