CLASS II ONGOING Device recall · Reported 14 Jan 2026 · FDA Enforcement Report
The Elecsys Anti-TSHR immunoassay is a three-step recalled
Roche Diagnostics Operations is recalling The Elecsys Anti-TSHR immunoassay is a three-step competitive immunoassay with, distributed nationwide in the US. The recall was initiated on 21 Nov 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1009-2026
- FDA event ID
- 98103
- Recalling firm
- Roche Diagnostics Operations, Indianapolis, IN
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 21 Nov 2025
- FDA classified
- 2 Jan 2026
- Reported
- 14 Jan 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Particulate Matter
- Device type
- Test Kits & Diagnostics
Timeline
| 21 Nov 2025 | Recall initiated by company | |
| 2 Jan 2026 | Classified by FDA | +42 days |
| 14 Jan 2026 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Issues identified: 1) Results vary by Instrument Platform; 2) Calibrator Lot-to-Lot Variability. Issues may lead to delayed or incorrect diagnosis, as well as delay of treatment or incorrect treatment.
Product as listed by the FDA
The Elecsys Anti-TSHR immunoassay is a three-step competitive immunoassay with streptavidin-coated microparticles and electrochemiluminescence detection. Patient serum sample is incubated with pretreatment reagent PT1 and the reconstituted PT2 containing a pre-formed immunocomplex of solubilized porcine TSH receptor (pTSHR) and biotinylated anti-pTSHR mouse monoclonal antibody.
Where it was distributed
US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, ME, MI, MN, NC, NJ, NM, NY, OH, PA, PR, TN, TX, UT, WA, WV.
Nationwide ArizonaCaliforniaFloridaGeorgiaIllinoisIndianaMaineMichiganMinnesotaNorth CarolinaNew JerseyNew MexicoNew YorkOhioPennsylvaniaPuerto RicoTennesseeTexasUtahWashington
Questions about this recall
Why was it recalled?
Issues identified: 1) Results vary by Instrument Platform; 2) Calibrator Lot-to-Lot Variability. Issues may lead to delayed or incorrect diagnosis, as well as delay of treatment or incorrect treatment.
Which lots are affected?
08496633190 lot 840183 exp. 02/26 08496633190 lot 874011 exp. 10/26 08496609190 lot 840177 exp. 05/26
Where was it sold?
US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, ME, MI, MN, NC, NJ, NM, NY, OH, PA, PR, TN, TX, UT, WA, WV.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Roche Diagnostics Operations
All 127| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | cobas pro integrated solutions recalled by Roche Diagnostics Operations | Roche Diagnostics Operations | Potency & Assay | Nationwide |
| Device | CLASS II | cobas HCYS, Homocysteine Enzymatic Assay, Material Number recalled | Roche Diagnostics Operations | Potency & Assay | 8 states |
| Device | CLASS II | Creatine Kinase (CK) used on cobas c 311/501/502 recalled | Roche Diagnostics Operations | Potency & Assay | Nationwide |
| Device | CLASS II | Tina-quant D-Dimer Test System recalled by Roche Diagnostics Operations | Roche Diagnostics Operations | Potency & Assay | Nationwide |
| Device | CLASS II | cobas infinity central lab, Material Number recalled | Roche Diagnostics Operations | Device Malfunction | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,663| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DxC 500 AU Clinical Chemistry Analyzers: AU w/out ISE recalled | Beckman Coulter | Device Malfunction | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules: DxC 500 AU Module w/ISE recalled | Beckman Coulter | Device Malfunction | Nationwide |
| Device | CLASS II | McKesson Consult hCG Urine Cassette recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | CLASS II | Systems/units containing affected recalled | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |