CLASS II Device recall · Reported 16 Jan 2013 · FDA Enforcement Report

Emit(R) 2000 Phenobarbital Assay (OSR4D229) recalled

Siemens Healthcare Diagnostics is recalling Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay, distributed in 18 states. The recall was initiated on 10 Dec 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesEmit(R) 2000 Phenobarbital Assay (OSR4D229) lots 1237 and 1264 and/or Syva(R) Emit(R) Phenobarbital Assay (4D019UL), lots E1, E2, and E3.
Quantity recalled1979 cartons

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0664-2013
FDA event ID
63892
Recalling firm
Siemens Healthcare Diagnostics, Newark, DE
Classification
Class II
Status
Terminated
Recall initiated
10 Dec 2012
FDA classified
10 Jan 2013
Reported
16 Jan 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

10 Dec 2012Recall initiated by company
10 Jan 2013Classified by FDA+31 days
16 Jan 2013Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Firm has confirmed higher imprecision with the recalled lots on Beckman Coulter AU(R) Systems and the Syva(R) Emit(R) 2000 Phenobarbital (4D019UL) lots E1, E2, and E3 as an application if the AU(R) System is used, the issue may be experienced. Firm has found that within-run imprecision of the recalled lots frequency of outliers that differ from the expected value by 20%. This is an expansion of

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay (4D019UL). Homogeneous enzyme immunoassay intended for use in the quantitative analysis of Phenobarbital in human serum or plasma.

Where it was distributed

Distributed in the states of CA, FL, IL, KS, MA, MI, MN, MS, NM, NY, NC, OH, OR, PA, TN, TX, UT, and WA.

CaliforniaFloridaIllinoisKansasMassachusettsMichiganMinnesotaMississippiNorth CarolinaNew MexicoNew YorkOhioOregonPennsylvaniaTennesseeTexasUtahWashington

Questions about this recall

Why was it recalled?

Firm has confirmed higher imprecision with the recalled lots on Beckman Coulter AU(R) Systems and the Syva(R) Emit(R) 2000 Phenobarbital (4D019UL) lots E1, E2, and E3 as an application if the AU(R) System is used, the issue may be experienced. Firm has found that within-run imprecision of the recalled lots frequency of outliers that differ from the expected value by 20%. This is an expansion of

Which lots are affected?

Emit(R) 2000 Phenobarbital Assay (OSR4D229) lots 1237 and 1264 and/or Syva(R) Emit(R) Phenobarbital Assay (4D019UL), lots E1, E2, and E3.

Where was it sold?

Distributed in the states of CA, FL, IL, KS, MA, MI, MN, MS, NM, NY, NC, OH, OR, PA, TN, TX, UT, and WA.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.