CLASS II ONGOING Device recall · Reported 4 Dec 2024 · FDA Enforcement Report
Essenz HLM (Heart-Lung Machine) recalled by LivaNova Deutschland
LivaNova Deutschl is recalling Essenz HLM (Heart-Lung Machine), distributed nationwide in the US. The recall was initiated on 18 Oct 2024 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0571-2025
- FDA event ID
- 95665
- Recalling firm
- LivaNova Deutschl, Munich
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 18 Oct 2024
- FDA classified
- 27 Nov 2024
- Reported
- 4 Dec 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Medical Software
Timeline
| 18 Oct 2024 | Recall initiated by company | |
| 27 Nov 2024 | Classified by FDA | +40 days |
| 4 Dec 2024 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby, then gas blender must be reactivated using user interface on blender, sensors deactivated until new case started; GUI reset can lead to Hypoxia and Hypoperfusion.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software
Where it was distributed
US Nationwide distribution in the states of CA, VA, GA, IN, NJ, FL, CO, MS, NV.
Nationwide CaliforniaColoradoFloridaGeorgiaIndianaMississippiNew JerseyNevadaVirginia
Questions about this recall
Why was it recalled?
Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby, then gas blender must be reactivated using user interface on blender, sensors deactivated until new case started; GUI reset can lead to Hypoxia and Hypoperfusion.
Which lots are affected?
UDI-DI: 04033817903062. Software Version: 1.5. Serial Numbers: 49PK00436, 49PK00108, 49PK00374, 49PK00479, 49PK00480, 49PK00481, 49PK00093, 49PK00091, 49PK00092, 49PK00163, 49PK00497, 49PK00498, 49PK00219, 49PK00223, 49PK00224, 49PK00225, 49PK00226, 49PK00241, 49PK00255, 49PK00292, 49PK00291, 49PK00293, 49PK00490, 49PK00487, 49PK00488, 49PK00278, 49PK00274, 49PK00277, 49PK00275, 49PK00276, 49PK00273, 49PK00457, 49PK00458, 49PK00459, 49PK00455, 49PK00456, 49PK00290, 49PK00288
Where was it sold?
US Nationwide distribution in the states of CA, VA, GA, IN, NJ, FL, CO, MS, NV.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from LivaNova Deutschl
All 5| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | 3T Heater Cooler System recalled by LivaNova Deutschland | LivaNova Deutschl | Other | Nationwide |
| Device | CLASS III | Sorin Centrifugal Pump System, CP5, Drive Unit recalled | LivaNova Deutschl | Device Malfunction | FL, NE, TX |
| Device | CLASS II | Control Panel recalled by LivaNova Deutschland | LivaNova Deutschl | Other | 13 states |
| Device | CLASS II | Control Panel recalled by LivaNova Deutschland | LivaNova Deutschl | Other | 13 states |
Other Medical Software recalls
All 932| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS III | Kit: Urinary Clean Catch 20/cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Device Malfunction | CA |
| Device | CLASS II | Spectral CT and Spectral CT 7500 RT recalled by Philips Medical Systems | Philips Medical Systems | Device Malfunction | Nationwide |
| Device | CLASS II | SmartPath to dStream for XR and 3.0T recalled by Philips North America | Philips North America | Device Malfunction | Nationwide |
| Device | CLASS II | Upgrade to MR recalled by Philips North America | Philips North America | Device Malfunction | Nationwide |
| Device | CLASS II | Positron Emission Tomography and Computed Tomography System recalled | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |