CLASS II Device recall · Reported 8 Jul 2020 · FDA Enforcement Report

Control Panel recalled by LivaNova Deutschland

LivaNova Deutschl is recalling Control Panel, distributed in 13 states. The recall was initiated on 26 Aug 2019 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Number: 60S14838 60S14839 60S14829 60S14724 60S14850 60S14858 60S14732 60S14737 60S14844 60S14846 60S14854 60S14855 60S14777 60S14778 60S14725 60S14727 60S14784 60S14785 60S14845 60S14774 60S14775 60S14786 60S14830 60S14776 60S14882
Quantity recalled24

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2440-2020
FDA event ID
85505
Recalling firm
LivaNova Deutschl, Munich, N/A
Classification
Class II
Status
Terminated
Recall initiated
26 Aug 2019
FDA classified
26 Jun 2020
Reported
8 Jul 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

26 Aug 2019Recall initiated by company
26 Jun 2020Classified by FDA+305 days
8 Jul 2020Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The SCP control Panel rotary knob used to control rotary speed of the pump in the SCP system used as a centrifugal pump during cardiopulmonary bypass may be difficult or impossible to rotate as a result of premature wear of the angle encoder. The loss of function of the angle encoder during bypass may lead to ischemia, stroke, organ damage, arterial tissue injury or reperfusion syndrome.

Product as listed by the FDA

Control Panel REF - 60-02-15: LivaNova SCP Pump Control Panel. REF 60-02-15, IPX2, RxOnly, UDI: 04033817900986

Where it was distributed

US: FL, GA, IN, KY, MI, MO, NY, OH, PA, SC, TN, TX, VA OUS: Canada - Uzbekistan - Turkey - Tunisia - Taiwan - South Africa - Serbia - Saudi Arabia - Russia - Poland - Mexico - Korea(Rep. of) - India - Hong Kong - Colombia - China - Brazil - Bahrain - Argentina - Algeria - Italy - Germany

FloridaGeorgiaIndianaKentuckyMichiganMissouriNew YorkOhioPennsylvaniaSouth CarolinaTennesseeTexasVirginia

Questions about this recall

Why was it recalled?

The SCP control Panel rotary knob used to control rotary speed of the pump in the SCP system used as a centrifugal pump during cardiopulmonary bypass may be difficult or impossible to rotate as a result of premature wear of the angle encoder. The loss of function of the angle encoder during bypass may lead to ischemia, stroke, organ damage, arterial tissue injury or reperfusion syndrome.

Which lots are affected?

Serial Number: 60S14838 60S14839 60S14829 60S14724 60S14850 60S14858 60S14732 60S14737 60S14844 60S14846 60S14854 60S14855 60S14777 60S14778 60S14725 60S14727 60S14784 60S14785 60S14845 60S14774 60S14775 60S14786 60S14830 60S14776 60S14882

Where was it sold?

US: FL, GA, IN, KY, MI, MO, NY, OH, PA, SC, TN, TX, VA OUS: Canada - Uzbekistan - Turkey - Tunisia - Taiwan - South Africa - Serbia - Saudi Arabia - Russia - Poland - Mexico - Korea(Rep. of) - India - Hong Kong - Colombia - China - Brazil - Bahrain - Argentina - Algeria - Italy - Germany

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 85505
ReportedClassRecallCompanyReasonWhere
CLASS IIControl Panel recalled by LivaNova DeutschlandLivaNova DeutschlOther13 states

More recalls from LivaNova Deutschl

All 5

Other Other Medical Devices recalls

All 13,908