CLASS II Device recall · Reported 24 Feb 2016 · FDA Enforcement Report

Evolis Microplate System recalled by Bio-Rad Laboratories

Bio-Rad Laboratories is recalling Evolis Microplate System, distributed nationwide in the US. The recall was initiated on 22 Dec 2015 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesThe EVOLIS Microplate System instrument is not lot controlled, however for traceability purposes each instrument is assigned a unique serial number. SERIAL NUMBERS: 9163700126, 9163700175, 9163700187, 9163700196, 9163700208, 9163700210, 9163700231, 9163700266, 9163700278, 9163700320, 9163700328, 9163700336, 9163700337, 9163700349, 9163700354, 9163700368, 9163700402, 9163700431, 9163700449, 9163700478, 9163700480, 9163700481, 9163700495, 9163700498, 9163700499, 9163700501, 9163700510, 9163700527, 9163700530, 9163700531, 9163700544, 9163700556, 9163700558, 9163700559, 9163700560, 9163700567, 9163700585, 9163700587, 9163700594, 9163700595, 9163700623, 9163700624, 9163700625, 9163700629, 9163700656, 9163700695, 9163700696, 9163700722, 9163700743, 9163700744, 9163700745, 9163700764, 9163700765, 9163700772, 9163700775, 9163700776, 9163700810, 9163700821, 9163700822, 9163700823, 9163700835, 9163700836, 9163700843, 9163700844, 9163700846, 9163700850, 9163700851, 9163700853, 9163700855, 9163700857, 9163700862, 9163700863, 9163700864, 9163700868, 9163700873, 9163700874, 9163700875, 9163700877, 9163700879, 9163700883, 9163700884, 9163700885, 9163700887, 9163700888, 9163700889, 9163700890, 9163700891, 9163700892, 9163700893, 9163700895, 9163700898, 9163700899, 9163700901, 9163700902, 9163700906, 9163700907, 9163700909, 9163700910, 9163700911, 9163700912, 9163700913, 9163700914, 9163700916, 9163700917, 9163700918, 9163700919, 9163700920, 9163700921, 9163700922, 9163700926, 9163700927, 9163700928, 9163700929, 9163700930, 9163700931, 9163700932, 9163700933, 9163700934, 9163700935, 9163700936, 9163700937, 9163700938, 9163700939, 9163700941, 9163700942, 9163700943, 9163700945, 9163700946, 9163700947, 9163700948, 9163700949, 9163700950, 9163700952, 9163700956, 9163700957, 9163700959, 9163700960, 9163700961, 9163700962, 9163700964, 9163700965, 9163700966, 9163700967, 9163700968, 9163700972, 9163700973, 9163700974, 9163700975, 9163700977, 9163700978, 9163700979, 9163700980, 9163700981, 9163700982, 9163700984, 9163700985, 9163700987, 9163700988, 9163700989, 9163700991, 9163700992, 9163700993, 9163700996, 9163700997, 9163700998, 9163700999, 9163701000, 9163701001, 9163701002, 9163701003, 9163701004, 9163701005, 9163701006, 9163701007, 9163701008, 9163701009, 9163701010, 9163701011, 9163701012, 9163701013, 9163701014, 9163701015, 9163701017, 9163701018, 9163701019, 9163701020, 9163701021, 9163701022, 9163701023, 9163701024, 9163701025, 9163701026, 9163701027, 9163701028, 9163701029, 9163701030, 9163701031, 9163701032, 9163701033, 9163701034, 9163701035, 9163701036, 9163701037, 9163701038, 9163701039, 9163701040, 9163701041, 9163701042, 9163701043, 9163701044, 9163701045, 9163701046, 9163701047, 9163701048, 9163701049, 9163701051, 9163701052, 9163701053, 9163701054, 9163701056, 9163701057, 9163701058, 9163701059, 9163701060, 9163701061, 9163701063, 9163700168R, 9163700172R, 9163700176R, 9163700202R, 9163700206R, 9163700209R, 9163700215R, 9163700221R, 9163700234R, 9163700236R, 9163700242R, 9163700246R, 9163700247R, 9163700276R, 9163700281R, 9163700282R, 9163700293R, 9163700294R, 9163700296R, 9163700326R, 9163700327R, 9163700330R, 9163700355R, 9163700359R, 9163700397R, 9163700398R, 9163700400R, 9163700432R, 9163700479R, 9163700494R, 9163700562R, 9163700563R, 9163700564R, 9163700565R, 9163700573R, 9163700617R, 9163700811R, 9163700834R, 9163700837R, 9163700878R, 9163700881R, 9163700886R, 9163700904R, 9163700908R, 9163700915R, 9163700925R, 9163700951R, 9163700958R, 9163700969R, 9163700971R, 9163700986R, and 916370726R.
Quantity recalled278 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0807-2016
FDA event ID
73030
Recalling firm
Bio-Rad Laboratories, Redmond, WA
Classification
Class II
Status
Terminated
Recall initiated
22 Dec 2015
FDA classified
12 Feb 2016
Reported
24 Feb 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

22 Dec 2015Recall initiated by company
12 Feb 2016Classified by FDA+52 days
24 Feb 2016Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The error "Washer Reagent Clean Fluid" level low alarmed, which prompted the customer to open the system drawer. When the system drawer was opened the customer noted the plate transport was still running and as a consequence the plate was pushed off the plate transporter.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

EVOLIS Microplate System, Catalog # 89601. Part number 89788 for the EVOLIS Operator's Manual. in vitro diagnostic Product Usage: EVOLIS Microplate System is a 4 plate fully integrated microplate processing system designed for use with multiple EIA assays.

Where it was distributed

US Nationwide Distribution including Puerto Rico.

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

The error "Washer Reagent Clean Fluid" level low alarmed, which prompted the customer to open the system drawer. When the system drawer was opened the customer noted the plate transport was still running and as a consequence the plate was pushed off the plate transporter.

Which lots are affected?

The EVOLIS Microplate System instrument is not lot controlled, however for traceability purposes each instrument is assigned a unique serial number. SERIAL NUMBERS: 9163700126, 9163700175, 9163700187, 9163700196, 9163700208, 9163700210, 9163700231, 9163700266, 9163700278, 9163700320, 9163700328, 9163700336, 9163700337, 9163700349, 9163700354, 9163700368, 9163700402, 9163700431, 9163700449, 9163700478, 9163700480, 9163700481, 9163700495, 9163700498, 9163700499, 9163700501, 9163700510, 9163700527, 9163700530, 9163700531, 9163700544, 9163700556, 9163700558, 9163700559, 9163700560, 9163700567, 9163700585, 9163700587, 9163700594, 9163700595, 9163700623, 9163700624, 9163700625, 9163700629, 9163700656, 9163700695, 9163700696, 9163700722, 9163700743, 9163700744, 9163700745, 9163700764, 9163700765, 9163700772, 9163700775, 9163700776, 9163700810, 9163700821, 9163700822, 9163700823, 9163700835, 9163700836, 9163700843, 9163700844, 9163700846, 9163700850, 9163700851, 9163700853, 9163700855, 9163700857, 9163700862, 9163700863, 9163700864, 9163700868, 9163700873, 9163700874, 9163700875, 9163700877, 9163700879, 9163700883, 9163700884, 9163700885, 9163700887, 9163700888, 9163700889, 9163700890, 9163700891, 9163700892, 9163700893, 9163700895, 9163700898, 9163700899, 9163700901, 9163700902, 9163700906, 9163700907, 9163700909, 9163700910, 9163700911, 9163700912, 9163700913, 9163700914, 9163700916, 9163700917, 9163700918, 9163700919, 9163700920, 9163700921, 9163700922, 9163700926, 9163700927, 9163700928, 9163700929, 9163700930, 9163700931, 9163700932, 9163700933, 9163700934, 9163700935, 9163700936, 9163700937, 9163700938, 9163700939, 9163700941, 9163700942, 9163700943, 9163700945, 9163700946, 9163700947, 9163700948, 9163700949, 9163700950, 9163700952, 9163700956, 9163700957, 9163700959, 9163700960, 9163700961, 9163700962, 9163700964, 9163700965, 9163700966, 9163700967, 9163700968, 9163700972, 9163700973, 9163700974, 9163700975, 9163700977, 9163700978, 9163700979, 9163700980, 9163700981, 9163700982, 9163700984, 9163700985, 9163700987, 9163700988, 9163700989, 9163700991, 9163700992, 9163700993, 9163700996, 9163700997, 9163700998, 9163700999, 9163701000, 9163701001, 9163701002, 9163701003, 9163701004, 9163701005, 9163701006, 9163701007, 9163701008, 9163701009, 9163701010, 9163701011, 9163701012, 9163701013, 9163701014, 9163701015, 9163701017, 9163701018, 9163701019, 9163701020, 9163701021, 9163701022, 9163701023, 9163701024, 9163701025, 9163701026, 9163701027, 9163701028, 9163701029, 9163701030, 9163701031, 9163701032, 9163701033, 9163701034, 9163701035, 9163701036, 9163701037, 9163701038, 9163701039, 9163701040, 9163701041, 9163701042, 9163701043, 9163701044, 9163701045, 9163701046, 9163701047, 9163701048, 9163701049, 9163701051, 9163701052, 9163701053, 9163701054, 9163701056, 9163701057, 9163701058, 9163701059, 9163701060, 9163701061, 9163701063, 9163700168R, 9163700172R, 9163700176R, 9163700202R, 9163700206R, 9163700209R, 9163700215R, 9163700221R, 9163700234R, 9163700236R, 9163700242R, 9163700246R, 9163700247R, 9163700276R, 9163700281R, 9163700282R, 9163700293R, 9163700294R, 9163700296R, 9163700326R, 9163700327R, 9163700330R, 9163700355R, 9163700359R, 9163700397R, 9163700398R, 9163700400R, 9163700432R, 9163700479R, 9163700494R, 9163700562R, 9163700563R, 9163700564R, 9163700565R, 9163700573R, 9163700617R, 9163700811R, 9163700834R, 9163700837R, 9163700878R, 9163700881R, 9163700886R, 9163700904R, 9163700908R, 9163700915R, 9163700925R, 9163700951R, 9163700958R, 9163700969R, 9163700971R, 9163700986R, and 916370726R.

Where was it sold?

US Nationwide Distribution including Puerto Rico.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.