CLASS II Device recall · Reported 17 Mar 2021 · FDA Enforcement Report

ExacTrac Dynamic (Medical Charged-Particle Radiation Therapy System) recalled

Brainlab is recalling ExacTrac Dynamic (Medical Charged-Particle Radiation Therapy System), distributed in 8 states. The recall was initiated on 1 Mar 2021 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel/catalogue numbers: 20910-06 ETD Implanted lic. UDI: The GTIN for ExacTrac Dynamic version 1.0 is 04056481142315. GMDN code: 40887. Software revisions of affected ExacTrac Dynamic software: versions 1.0.0, 1.0.1, 1.0.2.
Quantity recalled73 ExacTrac Dynamic systems (versions 1.0.0, 1.0.1, 1.0.2) have been manufactured and distributed

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1185-2021
FDA event ID
87395
Recalling firm
Brainlab, Munich
Classification
Class II
Status
Terminated
Recall initiated
1 Mar 2021
FDA classified
8 Mar 2021
Reported
17 Mar 2021
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Device Malfunction

Timeline

1 Mar 2021Recall initiated by company
8 Mar 2021Classified by FDA+7 days
17 Mar 2021Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

In case of a failed automatic marker detection, a software error causes parts of the display to incorrectly behave as if the current patient position is within predefined tolerances and may allow the user to proceed to treatment despite potentially exceeding shift values.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

ExacTrac Dynamic (Medical Charged-Particle Radiation Therapy System).

Where it was distributed

USA: (CA, NJ, AZ, AR, NY, IN, OR, NC), Australia, Austria, Belgium, Denmark, France, Germany, Hong Kong, Italy, Slovakia, Spain, Thailand, Turkey, United Kingdom, United Arab Emirates.

ArkansasArizonaCaliforniaIndianaNorth CarolinaNew JerseyNew YorkOregon

Questions about this recall

Why was it recalled?

In case of a failed automatic marker detection, a software error causes parts of the display to incorrectly behave as if the current patient position is within predefined tolerances and may allow the user to proceed to treatment despite potentially exceeding shift values.

Which lots are affected?

Model/catalogue numbers: 20910-06 ETD Implanted lic. UDI: The GTIN for ExacTrac Dynamic version 1.0 is 04056481142315. GMDN code: 40887. Software revisions of affected ExacTrac Dynamic software: versions 1.0.0, 1.0.1, 1.0.2.

Where was it sold?

USA: (CA, NJ, AZ, AR, NY, IN, OR, NC), Australia, Austria, Belgium, Denmark, France, Germany, Hong Kong, Italy, Slovakia, Spain, Thailand, Turkey, United Kingdom, United Arab Emirates.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Brainlab

All 39

Other Other Medical Devices recalls

All 13,908