CLASS II Device recall · Reported 27 Dec 2017 · FDA Enforcement Report

ExploR Radial Head Replacement System recalled by Zimmer Biomet

Zimmer Biomet is recalling ExploR Radial Head Replacement System. The recall was initiated on 16 Jun 2017 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPart Lot UDI Number 11-210061 243610 and 202900; (01)00880304210448(17)260817(10)243610

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0269-2018
FDA event ID
78705
Recalling firm
Zimmer Biomet, Warsaw, IN
Classification
Class II
Status
Terminated
Recall initiated
16 Jun 2017
FDA classified
20 Dec 2017
Reported
27 Dec 2017
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Packaging Defects

Timeline

16 Jun 2017Recall initiated by company
20 Dec 2017Classified by FDA+187 days
27 Dec 2017Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Zimmer Biomet is conducting a medical device recall for the ExploR 6x24mm Implantable Stem with Screw due to a complaint alleging that the Modular Radial HD lock screw was not included in the package for the stem.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

ExploR Radial Head Replacement System. PN: 11-210061 LN: 243610. The stem is placed inside a polyurethane bag. The bagged component will then be placed in an inner PETG tray. The screw will be placed inside a separate polybag, and placed in the PETG tray with foam pieces added to the top and bottom of the tray for protection. The inner tray and the contents within will then have a Tyvek¿ lid heat sealed to the tray. The sealed inner tray will be fit into an outer PETG tray and a Tyvek¿ lid heat sealed to the tray. The entire sealed unit is placed in a SBS carton that will be shrink wrapped for protection.

Where it was distributed

US, Canada, OUS.

Questions about this recall

Why was it recalled?

Zimmer Biomet is conducting a medical device recall for the ExploR 6x24mm Implantable Stem with Screw due to a complaint alleging that the Modular Radial HD lock screw was not included in the package for the stem.

Which lots are affected?

Part Lot UDI Number 11-210061 243610 and 202900; (01)00880304210448(17)260817(10)243610

Where was it sold?

US, Canada, OUS.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Zimmer Biomet

All 873

Other Masks, Gloves & PPE recalls

All 450