CLASS II Device recall · Reported 18 Jul 2018 · FDA Enforcement Report

Fdi Fort Defiance Industries P2131 Automated Field Steam recalled

Fort Defiance Industries is recalling Fdi Fort Defiance Industries P2131 Automated Field Steam Sterilizer, Nsn: 6530-01-641-4641, distributed nationwide in the US. The recall was initiated on 24 May 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Numbers: AFS-0001 through AFS-0199
Quantity recalled199 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2411-2018
FDA event ID
80293
Recalling firm
Fort Defiance Industries, Loudon, TN
Classification
Class II
Status
Terminated
Recall initiated
24 May 2018
FDA classified
10 Jul 2018
Reported
18 Jul 2018
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Other Reasons

Timeline

24 May 2018Recall initiated by company
10 Jul 2018Classified by FDA+47 days
18 Jul 2018Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

During long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the plumbing of the P2131 device produces deposits in the water system that can potentially interfere with the rear sight gauge and / or the water liquid level switch.

Product as listed by the FDA

FDI FORT DEFIANCE INDUSTRIES P2131 AUTOMATED FIELD STEAM STERILIZER, NSN: 6530-01-641-4641 Product Usage: The Fort Defiance Industries Automated Field Steam Sterilizer Model P2131 is designed for sterilization of porous and non-porous, heat- and moisture-stable materials (e.g., surgical instruments and textiles) used in healthcare facilities. The P2131 sterilizer is a transportable device designed for use in a variety of austere environments.

Where it was distributed

US Nationwide Distribution in the states of PA. VA. TX

Nationwide PennsylvaniaTexasVirginia

Questions about this recall

Why was it recalled?

During long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the plumbing of the P2131 device produces deposits in the water system that can potentially interfere with the rear sight gauge and / or the water liquid level switch.

Which lots are affected?

Serial Numbers: AFS-0001 through AFS-0199

Where was it sold?

US Nationwide Distribution in the states of PA. VA. TX

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Fort Defiance Industries

All 3

Other Surgical Instruments recalls

All 3,891