CLASS II Device recall · Reported 3 Sep 2014 · FDA Enforcement Report

Fluid Management System is designed to provide irrigation recalled

Thermedx is recalling The Thermedx Fluid Management System is designed to provide irrigation fluids to irrigate, distributed nationwide in the US. The recall was initiated on 10 Feb 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel P4000 Serial numbers affected: 20110004, 20110007, 20110015, 20110025, 20130001-20130032, 20140001-20140005
Quantity recalled21

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2555-2014
FDA event ID
68986
Recalling firm
Thermedx, Solon, OH
Classification
Class II
Status
Terminated
Recall initiated
10 Feb 2014
FDA classified
27 Aug 2014
Reported
3 Sep 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

10 Feb 2014Recall initiated by company
27 Aug 2014Classified by FDA+198 days
3 Sep 2014Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A vendor changed the manufacturing process of a component used in the canister ring that may affect the performance of the ring. The deficit display may be higher than the actual deficit when the fluid collection canister is fully loaded.

Product as listed by the FDA

The Thermedx Fluid Management System is designed to provide irrigation fluids to irrigate, distend, and debride while warming the fluid to assist in the prevention of peri-operative hypothermia. The fluid is warmed via an internal infrared mechanism. Additionally the system will hold suction canisters for the collection of used irrigant.

Where it was distributed

Nationwide Distribution including HI, MS, NY, OH, PA, VA, WA, WV, and IL.

Nationwide IllinoisMississippiNew YorkOhioPennsylvaniaVirginiaWashingtonWest Virginia

Questions about this recall

Is Fluid Management System is designed to provide irrigation fluids to irrigate, distend recalled?

Yes. Thermedx recalled Fluid Management System is designed to provide irrigation fluids to irrigate, distend on 10 Feb 2014. The recall is Class II and its status is Terminated. Recall number Z-2555-2014.

Why was it recalled?

A vendor changed the manufacturing process of a component used in the canister ring that may affect the performance of the ring. The deficit display may be higher than the actual deficit when the fluid collection canister is fully loaded.

Which lots are affected?

Model P4000 Serial numbers affected: 20110004, 20110007, 20110015, 20110025, 20130001-20130032, 20140001-20140005

Where was it sold?

Nationwide Distribution including HI, MS, NY, OH, PA, VA, WA, WV, and IL.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Thermedx

All 2
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIFluid Management System P4000 recalled by ThermedxThermedxDevice Malfunction11 states

Other Other Medical Devices recalls

All 13,778