CLASS II ONGOING Device recall · Reported 11 Feb 2026 · FDA Enforcement Report

Focalyx Fusion recalled by Focalyx Technologies

Focalyx Technologies is recalling Focalyx Fusion, distributed nationwide in the US. The recall was initiated on 23 Dec 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 00860003552001. Software Versions: 1.3, 1.5, 1.6. Serial Numbers: IF-FF-B-1-22080504, IF-FF-B-1-TPAA998118, IF-FF-B-1-351238, IF-FF-B-1-TPAB081335, IF-FF-B-1-22080508, IF-FF-B-1-TPAB200331, IF-FF-B-1-TPAB200333, IF-FF-B-1-TPAC505383, IF-FF-B-1-TPAB200348, IF-FF-B-1-TPAB540364, IF-FF-B-1-S5ARQL009253NYZ, IF-FF-B-1-SCARQL0014096YZ, IF-FF-B-1-TPAB532163, IF-FF-B-1-SCARQL00136033KK, IF-FF-B-1-TPAB540359, IF-FF-B-1-TPAB081326, IF-FF-B-1-22080505, IF-FF-B-1-TPAB081324
Quantity recalled18

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1243-2026
FDA event ID
98137
Recalling firm
Focalyx Technologies, Hialeah, FL
Classification
Class II
Status
Ongoing
Recall initiated
23 Dec 2025
FDA classified
3 Feb 2026
Reported
11 Feb 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

23 Dec 2025Recall initiated by company
3 Feb 2026Classified by FDA+42 days
11 Feb 2026Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, which may cause patient harm during biopsy or cause treatment/therapy delays, so firm is recommending stop use until device is verified/validated with Windows 11.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Focalyx Fusion

Where it was distributed

Worldwide - US Nationwide distribution in the states of NY, MA, FL and the countries of ES, VE, DO.

Nationwide FloridaMassachusettsNew York

Questions about this recall

Why was it recalled?

Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, which may cause patient harm during biopsy or cause treatment/therapy delays, so firm is recommending stop use until device is verified/validated with Windows 11.

Which lots are affected?

UDI-DI: 00860003552001. Software Versions: 1.3, 1.5, 1.6. Serial Numbers: IF-FF-B-1-22080504, IF-FF-B-1-TPAA998118, IF-FF-B-1-351238, IF-FF-B-1-TPAB081335, IF-FF-B-1-22080508, IF-FF-B-1-TPAB200331, IF-FF-B-1-TPAB200333, IF-FF-B-1-TPAC505383, IF-FF-B-1-TPAB200348, IF-FF-B-1-TPAB540364, IF-FF-B-1-S5ARQL009253NYZ, IF-FF-B-1-SCARQL0014096YZ, IF-FF-B-1-TPAB532163, IF-FF-B-1-SCARQL00136033KK, IF-FF-B-1-TPAB540359, IF-FF-B-1-TPAB081326, IF-FF-B-1-22080505, IF-FF-B-1-TPAB081324

Where was it sold?

Worldwide - US Nationwide distribution in the states of NY, MA, FL and the countries of ES, VE, DO.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.