CLASS II ONGOING Device recall · Reported 19 Nov 2025 · FDA Enforcement Report

Food Panel 3 Specific Allergen recalled

Siemens Healthcare Diagnostics is recalling Food Panel 3 Specific Allergen, distributed nationwide in the US. The recall was initiated on 6 Oct 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSiemens Material Number (SMN): 10385891; Model/Catalog: FP3L4; UDI-DI: 00630414960418; Lot number: 629;
Quantity recalled28 units (OUS only)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0493-2026
FDA event ID
97819
Recalling firm
Siemens Healthcare Diagnostics, East Walpole, MA
Classification
Class II
Status
Ongoing
Recall initiated
6 Oct 2025
FDA classified
13 Nov 2025
Reported
19 Nov 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Labeling Errors

Timeline

6 Oct 2025Recall initiated by company
13 Nov 2025Classified by FDA+38 days
19 Nov 2025Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.

Product as listed by the FDA

Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;

Where it was distributed

Worldwide - US Nationwide distribution in the state of CA and the countries of Argentina, Australia, Austria, Belgium, Bosnia Herzeg. , Brazil, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Israel, Italy, Latvia, Macedonia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Portugal, Romania, Russian Fed., Serbia, Slovakia, Spain, Sweden, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay.

Nationwide California

Questions about this recall

Why was it recalled?

The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.

Which lots are affected?

Siemens Material Number (SMN): 10385891; Model/Catalog: FP3L4; UDI-DI: 00630414960418; Lot number: 629;

Where was it sold?

Worldwide - US Nationwide distribution in the state of CA and the countries of Argentina, Australia, Austria, Belgium, Bosnia Herzeg. , Brazil, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Israel, Italy, Latvia, Macedonia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Portugal, Romania, Russian Fed., Serbia, Slovakia, Spain, Sweden, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.