CLASS II ONGOING Device recall · Reported 11 Dec 2024 · FDA Enforcement Report

Fusion Pro recalled by Epilog Laser

Epilog Laser is recalling Fusion Pro 24, distributed nationwide in the US. The recall was initiated on 8 Aug 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel 17000 Product Report Accession Number: 2021046-000
Quantity recalled751

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0524-2025
FDA event ID
95757
Recalling firm
Epilog Laser, Golden, CO
Classification
Class II
Status
Ongoing
Recall initiated
8 Aug 2024
FDA classified
3 Dec 2024
Reported
11 Dec 2024
Type
FDA Mandated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

8 Aug 2024Recall initiated by company
3 Dec 2024Classified by FDA+117 days
11 Dec 2024Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

a. A small number of Fusion Pro 24 units were released with an incorrectly placed Interlock device. As a result, the system can be operated with the front-loading door in an open position. b. Fails to comply with Class 2 designation, indicated in Part 6 of Laser Product Report. An incorrectly placed Interlock device can allow Class 4 laser radiation fields during operation.

Product as listed by the FDA

Fusion Pro 24, Model 17000

Where it was distributed

US Nationwide Distribution

Nationwide

Questions about this recall

Why was it recalled?

a. A small number of Fusion Pro 24 units were released with an incorrectly placed Interlock device. As a result, the system can be operated with the front-loading door in an open position. b. Fails to comply with Class 2 designation, indicated in Part 6 of Laser Product Report. An incorrectly placed Interlock device can allow Class 4 laser radiation fields during operation.

Which lots are affected?

Model 17000 Product Report Accession Number: 2021046-000

Where was it sold?

US Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other Other Medical Devices recalls