CLASS II Device recall · Reported 28 Dec 2016 · FDA Enforcement Report
GAM Kirschner Wire, Sterile R, T2 K-Wire, Sterile R recalled
Stryker Howmedica Osteonics is recalling GAM Kirschner Wire, Sterile R, T2 K-Wire, Sterile R and T2 K-Wire Recon, Stryker Trauma, distributed nationwide in the US. The recall was initiated on 9 Nov 2016 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0880-2017
- FDA event ID
- 75792
- Recalling firm
- Stryker Howmedica Osteonics, Mahwah, NJ
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 9 Nov 2016
- FDA classified
- 22 Dec 2016
- Reported
- 28 Dec 2016
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Sterility
- Device type
- Other Medical Devices
Timeline
| 9 Nov 2016 | Recall initiated by company | |
| 22 Dec 2016 | Classified by FDA | +43 days |
| 28 Dec 2016 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
A review of packaging revealed the seal integrity of the pouch may be compromised. More specifically, there is a potential that the sterile pouch is not sealed at one end due to a manufacturing error.
A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.
Product as listed by the FDA
GAM Kirschner Wire, Sterile R, T2 K-Wire, Sterile R and T2 K-Wire Recon, Stryker Trauma GmbH, Prof.-Kuntscher-Strasse 1-5, Schonkirchen, Germany 24232; T2 K-Wire, Sterile R Intended for the temporary stabilization of bone segments or fragments
Where it was distributed
Worldwide Distribution - US (nationwide) Internationally to AU, CA, CL, CH, ES, FR, GB, SE, JP, FR, ZA, and NL.
Questions about this recall
Why was it recalled?
A review of packaging revealed the seal integrity of the pouch may be compromised. More specifically, there is a potential that the sterile pouch is not sealed at one end due to a manufacturing error.
Which lots are affected?
GAM Kirschner Wire - Manufacturing Part Number 12106450S, Lot #'s K0800BB, K0800BC, K0800BD, K0800BE, K0800BF, K0800C0, K081720, K081721, K0810872, K081723, K081727, K082C8B, K084F89, K084F98, K08683D, K08683E, K086841, K08820B, K089AF5, K08E1E1, K0911F4, K09379B, K096A26, K096A2A, K09A2C, K098213, K098215, K099AA0, K09AD33, K09BA4F, K09D564, K09F026, K0A1EF8, K0A1EFA, KoA1EFB, K0A38FB, K0A63AB, K0A63AC and K0A7BC1 T2 K-Wire - Manufacturing Part Number 18060050S, Lot #'s K084FBF, K0920A4, K0A1EFD, K093747, K0937C4, K09BA53, K09BA54, K09BA55 and K09BA56 T2 K-Wire Recon - Manufacturing Part Number 18063030S, Lot #'s K086847, K08E1E9 and K099AA8
Where was it sold?
Worldwide Distribution - US (nationwide) Internationally to AU, CA, CL, CH, ES, FR, GB, SE, JP, FR, ZA, and NL.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Stryker Howmedica Osteonics
All 86| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Hoffmann LRF Safety Clip (Bone Transport Strut) recalled | Stryker Howmedica Osteonics | Other | Nationwide |
| Device | CLASS II | Reunion TSA Peg Alignment Sound recalled by Stryker Howmedica Osteonics | Stryker Howmedica Osteonics | Other | Nationwide |
| Device | CLASS II | LFIT Anatomic V40 Femoral Head, Low Friction Ion Treatment, Sterile recalled | Stryker Howmedica Osteonics | Device Malfunction | Nationwide |
| Device | CLASS II | Stryker Orthopaedics Universal Acetabular Cup Positioner/Impactor recalled | Stryker Howmedica Osteonics | Other | NJ |
| Device | CLASS II | Stryker Orthopaedics Patella Assembly Instrument recalled | Stryker Howmedica Osteonics | Device Malfunction | Nationwide |
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |