CLASS II Device recall · Reported 16 Nov 2016 · FDA Enforcement Report

LFIT Anatomic V40 Femoral Head, Low Friction Ion Treatment, Sterile recalled

Stryker Howmedica Osteonics is recalling LFIT Anatomic V40 Femoral Head, Low Friction Ion Treatment, Sterile, distributed nationwide in the US. The recall was initiated on 29 Aug 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog #6260-9-236 - Head Diameter 36 mm, Offset +5, including all lots manufactured from 1/102 - 7/1/10; Catalog #6260-9-240 - Head Diameter 40 mm, Offset +4, including all lots manufactured from 1/1/06 - 3/4/11; Catalog #6260-9-244 - Head Diameter 44 mm, Offset +4, including all lots manufactured from 1/1/06 - 3/4/11; Catalog #6260-9-340 - Head Diameter 40 mm, Offset +8, including all lots manufactured from 1/1/06 - 3/4/11; Catalog #6260-9-440 - Head Diameter 40 mm, Offset +12, including all lots manufactured from 1/1/06 - 3/4/11; Catalog #6260-9-344 - Head Diameter 44 mm, Offset +8, including all lots manufactured from 1/1/07 - 3/4/11 and Catalog #6260-9-444 - Head Diameter 44 mm, Offset +12, including all lots manufactured from 1/1/06 - 3/4/11.
Quantity recalled42,519 units (total Catalog numbers)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0378-2017
FDA event ID
75246
Recalling firm
Stryker Howmedica Osteonics, Mahwah, NJ
Classification
Class II
Status
Terminated
Recall initiated
29 Aug 2016
FDA classified
9 Nov 2016
Reported
16 Nov 2016
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Device Malfunction

Timeline

29 Aug 2016Recall initiated by company
9 Nov 2016Classified by FDA+72 days
16 Nov 2016Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Stryker received several complaints describing incidence of harm secondary to taper lock failure for specific lots of numerous catalog numbers of LFIT Anatomic CoCr V40 Femoral Heads.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

LFIT Anatomic V40 Femoral Head, Low Friction Ion Treatment, Sterile, 36 mm, REF 6260-9-236; Modular components designed to be locked onto a femoral hip stem trunnion during surgery for total hip replacement.

Where it was distributed

US Nationwide and Internationally

Nationwide

Questions about this recall

Why was it recalled?

Stryker received several complaints describing incidence of harm secondary to taper lock failure for specific lots of numerous catalog numbers of LFIT Anatomic CoCr V40 Femoral Heads.

Which lots are affected?

Catalog #6260-9-236 - Head Diameter 36 mm, Offset +5, including all lots manufactured from 1/102 - 7/1/10; Catalog #6260-9-240 - Head Diameter 40 mm, Offset +4, including all lots manufactured from 1/1/06 - 3/4/11; Catalog #6260-9-244 - Head Diameter 44 mm, Offset +4, including all lots manufactured from 1/1/06 - 3/4/11; Catalog #6260-9-340 - Head Diameter 40 mm, Offset +8, including all lots manufactured from 1/1/06 - 3/4/11; Catalog #6260-9-440 - Head Diameter 40 mm, Offset +12, including all lots manufactured from 1/1/06 - 3/4/11; Catalog #6260-9-344 - Head Diameter 44 mm, Offset +8, including all lots manufactured from 1/1/07 - 3/4/11 and Catalog #6260-9-444 - Head Diameter 44 mm, Offset +12, including all lots manufactured from 1/1/06 - 3/4/11.

Where was it sold?

US Nationwide and Internationally

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Stryker Howmedica Osteonics

All 86

Other Hip & Knee Implants recalls

All 3,097