CLASS II Device recall · Reported 25 Apr 2018 · FDA Enforcement Report
GE Healthcare Lunar: a) DPX NT recalled by GE Medical Systems Ultrasound
GE Medical Systems Ultrasound & Primary Care Diagnostics, LL is recalling GE Healthcare Lunar: a) DPX NT, distributed nationwide in the US. The recall was initiated on 16 Mar 2018 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1394-2018
- FDA event ID
- 79710
- Recalling firm
- GE Medical Systems Ultrasound & Primary Care Diagnostics, LL, Madison, WI
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 16 Mar 2018
- FDA classified
- 17 Apr 2018
- Reported
- 25 Apr 2018
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Spinal Implants
Timeline
| 16 Mar 2018 | Recall initiated by company | |
| 17 Apr 2018 | Classified by FDA | +32 days |
| 25 Apr 2018 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS.
Product as listed by the FDA
GE Healthcare Lunar: a) DPX NT, Model Numbers: LU8230, LU40338, LU42357, LU42369 b) DPX MD+, Model Numbers: LU8230, LU40338, LU40352 Provides an estimate of BMD (Bone Marrow Density) at the lumbar spine and proximal femur regions. This BMD value can then be compared to a reference population at the sole discretion of the physician.
Where it was distributed
Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Lebanon, Malaysia, Malta, Mexico, Morocco, Norway, Poland, Portugal , Romania, Russia, Saudi Arabia, Spain, Switzerland, Taiwan, Tunisia, Turkey, United Arab, United Kingdom and Venezuela.
Questions about this recall
Why was it recalled?
Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS.
Which lots are affected?
a) DPX NT, Model Numbers: LU8230, LU40338, LU42357, LU42369 b) DPX MD+, Model Numbers: LU8230, LU40338, LU40352
Where was it sold?
Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Lebanon, Malaysia, Malta, Mexico, Morocco, Norway, Poland, Portugal , Romania, Russia, Saudi Arabia, Spain, Switzerland, Taiwan, Tunisia, Turkey, United Arab, United Kingdom and Venezuela.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
2 other products · FDA event 79710| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | GE Healthcare Lunar: a) DPX Duo recalled | GE Medical Systems Ultrasound & Primary Care Diagnostics, LL | Other | Nationwide | |
| CLASS II | GE Healthcare: a) Prodigy recalled by GE Medical Systems Ultrasound | GE Medical Systems Ultrasound & Primary Care Diagnostics, LL | Other | Nationwide |
More recalls from GE Medical Systems Ultrasound & Primary Care Diagnostics, LL
All 12Other Spinal Implants recalls
All 1,157| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Neuro; Neuro; Crani; Ortho Spine Pack; Bmhm Neuro Pack recalled | Avid Medical | Sterility | Nationwide |
| Device | CLASS II | Halyard medical convenience kits labeled as: ENT recalled | Avid Medical | Sterility | Nationwide |
| Device | CLASS II | Halyard medical convenience kits labeled as: A-P PACK recalled | Avid Medical | Sterility | Nationwide |
| Device | CLASS II | Medline medical convenience kits, labeled as recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline medical convenience kits, labeled as recalled | Medline Industries | Other | Nationwide |