CLASS II ONGOING Device recall · Reported 18 Mar 2026 · FDA Enforcement Report
GEM Premier 5000 PAK recalled by Instrumentation Laboratory
Instrumentation Laboratory is recalling GEM Premier 5000 PAK, distributed nationwide in the US. The recall was initiated on 20 Jan 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1524-2026
- FDA event ID
- 98349
- Recalling firm
- Instrumentation Laboratory, Bedford, MA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 20 Jan 2026
- FDA classified
- 10 Mar 2026
- Reported
- 18 Mar 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 20 Jan 2026 | Recall initiated by company | |
| 10 Mar 2026 | Classified by FDA | +49 days |
| 18 Mar 2026 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Product as listed by the FDA
GEM Premier 5000 PAK, Part No. 00055360004.
Where it was distributed
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Bulgaria, Brazil, Belarus, Switzerland, Canada, Chile, China, Colombia, Czech Republic, Germany, Denmark, Dominican Republic, Ecuador, Estonia, Spain, Finland, France, United Kingdom, Georgia, Greece, Guatemala, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, South Korea, Kazakhstan, Lebanon, Lithuania, Luxembourg, Latvia, Macau, Mexico, Malaysia, Netherlands, New Zealand, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, San Marino, South Africa, Thailand, Tunisia, Turkey, Uruguay, Uzbekistan & Vietnam.
Questions about this recall
Is GEM Premier 5000 PAK recalled?
Yes. Instrumentation Laboratory recalled GEM Premier 5000 PAK on 20 Jan 2026. The recall is Class II and its status is Ongoing. Recall number Z-1524-2026.
Why was it recalled?
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Which lots are affected?
Part number: 00055360004; UDI: 08426950899662; Lot No. 251113U, 251201AI, 251202Q, 251223E, 251223P, 251229X.
Where was it sold?
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Bulgaria, Brazil, Belarus, Switzerland, Canada, Chile, China, Colombia, Czech Republic, Germany, Denmark, Dominican Republic, Ecuador, Estonia, Spain, Finland, France, United Kingdom, Georgia, Greece, Guatemala, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, South Korea, Kazakhstan, Lebanon, Lithuania, Luxembourg, Latvia, Macau, Mexico, Malaysia, Netherlands, New Zealand, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, San Marino, South Africa, Thailand, Tunisia, Turkey, Uruguay, Uzbekistan & Vietnam.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Instrumentation Laboratory
All 77| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | The GEM Premier 7000 with iQM3 is a portable critical care recalled | Instrumentation Laboratory | Device Malfunction | Nationwide |
| Device | CLASS II | GEM Premier 5000 PAK Part Numbers recalled by Instrumentation Laboratory | Instrumentation Laboratory | Device Malfunction | Nationwide |
| Device | CLASS II | GEM Premier recalled by Instrumentation Laboratory | Instrumentation Laboratory | Other | Nationwide |
| Device | CLASS II | GEM Premier 5000 with IQM2 recalled by Instrumentation Laboratory | Instrumentation Laboratory | Other | Nationwide |
| Device | CLASS II | GEM Premier 7000 with iQM3 recalled by Instrumentation Laboratory | Instrumentation Laboratory | Other | Nationwide |
Other Other Medical Devices recalls
All 13,771| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits: Blood Draw Kit recalled by Medline Industries | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ortho Aspiration Kit recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep DYKA1054F recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Arterial Line Insertion Kit recalled | Medline Industries | Sterility | Nationwide |