CLASS II Device recall · Reported 24 Jun 2020 · FDA Enforcement Report

Giraffe Incubator with installed Servo Oxygen module recalled

GE Healthcare is recalling Giraffe Incubator with installed Servo Oxygen module, distributed nationwide in the US. The recall was initiated on 28 Apr 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesGiraffe Incubator: serial numbers HDGA through HDGQ; Servo Oxygen Board spare part kit: 6600-0233-850; Servo Oxygen upgrade kits: 6600-0678-800, 6600-0678-801, 6600-0678-802, 6600-0678-803, 66000849-800, 6600-0849-801, 6600-0849-802, 6600-0849-803, 6600-0866-800, 6600-0866-801, 6600-0867800, 6600-0867-801, M1142293 or M1142295.
Quantity recalled5,883 units total

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2236-2020
FDA event ID
85643
Recalling firm
GE Healthcare, Waukesha, WI
Classification
Class II
Status
Terminated
Recall initiated
28 Apr 2020
FDA classified
15 Jun 2020
Reported
24 Jun 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

28 Apr 2020Recall initiated by company
15 Jun 2020Classified by FDA+48 days
24 Jun 2020Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Potential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed.

Product as listed by the FDA

Giraffe Incubator with installed Servo Oxygen module. Infant incubator.

Where it was distributed

Worldwide Distribution. US nationwide including US Virgin Islands, Australia, Austria, Bahrain, Belgium, Bermuda, Bulgaria, Burundi, Canada, Chile, China, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guernsey, Hong Kong, India, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Republic of Korea, Kuwait, Latam, Latvia, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, OMAN, State of Palestine, Panama, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Yemen

Nationwide

Questions about this recall

Why was it recalled?

Potential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed.

Which lots are affected?

Giraffe Incubator: serial numbers HDGA through HDGQ; Servo Oxygen Board spare part kit: 6600-0233-850; Servo Oxygen upgrade kits: 6600-0678-800, 6600-0678-801, 6600-0678-802, 6600-0678-803, 66000849-800, 6600-0849-801, 6600-0849-802, 6600-0849-803, 6600-0866-800, 6600-0866-801, 6600-0867800, 6600-0867-801, M1142293 or M1142295.

Where was it sold?

Worldwide Distribution. US nationwide including US Virgin Islands, Australia, Austria, Bahrain, Belgium, Bermuda, Bulgaria, Burundi, Canada, Chile, China, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guernsey, Hong Kong, India, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Republic of Korea, Kuwait, Latam, Latvia, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, OMAN, State of Palestine, Panama, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Yemen

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 85643
ReportedClassRecallCompanyReasonWhere
CLASS IIGiraffe OmniBed with installed Servo Oxygen module recalledGE HealthcareOtherNationwide

More recalls from GE Healthcare

All 512

Other Other Medical Devices recalls

All 13,778