CLASS II ONGOING Device recall · Reported 21 Oct 2026 · FDA Enforcement Report

Grafton" DBM and Grafton Plus" DBM contain demineralized recalled

Medtronic Sofamor Danek USA is recalling Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with, distributed nationwide in the US. The recall was initiated on 12 Aug 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 00643169123021. Serial Numbers: A62881-034 through A62881-051, specifically: A62881-034, A62881-035, A62881-036, A62881-037, A62881-038, A62881-039, A62881-040, A62881-041, A62881-042, A62881-043, A62881-044, A62881-045, A62881-046, A62881-047, A62881-048, A62881-049, A62881-050, A62881-051
Quantity recalled18

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0137-2027
FDA event ID
99701
Recalling firm
Medtronic Sofamor Danek USA, Memphis, TN
Classification
Class II
Status
Ongoing
Recall initiated
12 Aug 2026
FDA classified
9 Oct 2026
Reported
21 Oct 2026
Type
Voluntary: Firm initiated

Timeline

12 Aug 2026Recall initiated by company
9 Oct 2026Classified by FDA+58 days
21 Oct 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured from the donor tissue. Donor bone tissue tested positive for Clostridium butyricum and while tissue did undergo gamma radiation, if implanted, there is a possibility of clostridium infection

Product as listed by the FDA

Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive to yield a demineralized bone matrix (DBM) allograft product having a particular physical form and/or handling property. Grafton DBM Gel and Grafton Plus DBM Paste are produced from a powder form of DBM, while Grafton DBM Flex, Putty, Matrix, Crunch" and Orthoblend are produced from a fiber form of DBM Crunch" also contains demineralized bone chips/cubes, while Grafton DBM Orthoblend contains non-demineralized cancellous bone chips in addition to DBM. Grafton DBM and Grafton Plus DBM are malleable and can be molded or cut into various sizes and shapes according to the intended implant site.

Where it was distributed

US Nationwide distribution in the states of OK, VA, CO, AR, FL, TX, MD, DE, IL, MI, KS, AL, GA.

Nationwide AlabamaArkansasColoradoDelawareFloridaGeorgiaIllinoisKansasMarylandMichiganTexasVirginia

Questions about this recall

Why was it recalled?

Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured from the donor tissue. Donor bone tissue tested positive for Clostridium butyricum and while tissue did undergo gamma radiation, if implanted, there is a possibility of clostridium infection

Which lots are affected?

UDI-DI: 00643169123021. Serial Numbers: A62881-034 through A62881-051, specifically: A62881-034, A62881-035, A62881-036, A62881-037, A62881-038, A62881-039, A62881-040, A62881-041, A62881-042, A62881-043, A62881-044, A62881-045, A62881-046, A62881-047, A62881-048, A62881-049, A62881-050, A62881-051

Where was it sold?

US Nationwide distribution in the states of OK, VA, CO, AR, FL, TX, MD, DE, IL, MI, KS, AL, GA.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 27 Oct 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 99701

This product is part of a recall event; the main page for the event is Grafton" DBM and Grafton Plus" DBM contain demineralized recalled.

More recalls from Medtronic Sofamor Danek USA

All 79

Other Hip & Knee Implants recalls

All 3,115