CLASS II ONGOING Device recall · Reported 21 Oct 2026 · FDA Enforcement Report
Grafton" DBM and Grafton Plus" DBM contain demineralized recalled
Medtronic Sofamor Danek USA is recalling Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with, distributed nationwide in the US. The recall was initiated on 12 Aug 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0138-2027
- FDA event ID
- 99701
- Recalling firm
- Medtronic Sofamor Danek USA, Memphis, TN
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 12 Aug 2026
- FDA classified
- 9 Oct 2026
- Reported
- 21 Oct 2026
- Type
- Voluntary: Firm initiated
- Reason category
- Other Bacterial Contamination
- Device type
- Hip & Knee Implants
Timeline
| 12 Aug 2026 | Recall initiated by company | |
| 9 Oct 2026 | Classified by FDA | +58 days |
| 21 Oct 2026 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured from the donor tissue. Donor bone tissue tested positive for Clostridium butyricum and while tissue did undergo gamma radiation, if implanted, there is a possibility of clostridium infection.
Product as listed by the FDA
Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive to yield a demineralized bone matrix (DBM) allograft product having a particular physical form and/or handling property. Grafton DBM Gel and Grafton Plus DBM Paste are produced from a powder form of DBM, while Grafton DBM Flex, Putty, Matrix, Crunch" and Orthoblend are produced from a fiber form of DBM Crunch" also contains demineralized bone chips/cubes, while Grafton DBM Orthoblend contains non-demineralized cancellous bone chips in addition to DBM. Grafton DBM and Grafton Plus DBM are malleable and can be molded or cut into various sizes and shapes according to the intended implant site.
Where it was distributed
US Nationwide distribution in the states of OK, VA, CO, AR, FL, TX, MD, DE, IL, MI, KS, AL, GA.
Nationwide AlabamaArkansasColoradoDelawareFloridaGeorgiaIllinoisKansasMarylandMichiganTexasVirginia
Questions about this recall
Why was it recalled?
Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured from the donor tissue. Donor bone tissue tested positive for Clostridium butyricum and while tissue did undergo gamma radiation, if implanted, there is a possibility of clostridium infection.
Which lots are affected?
UDI-DI: 00643169122932. Serial Numbers: A62881-014 through A62881-033, specifically: A62881-014, A62881-015, A62881-016, A62881-017, A62881-018, A62881-019, A62881-020, A62881-021, A62881-022, A62881-023, A62881-024, A62881-025, A62881-026, A62881-027, A62881-028, A62881-029, A62881-030, A62881-031, A62881-032, A62881-033
Where was it sold?
US Nationwide distribution in the states of OK, VA, CO, AR, FL, TX, MD, DE, IL, MI, KS, AL, GA.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 27 Oct 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
1 other product · FDA event 99701| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Grafton" DBM and Grafton Plus" DBM contain demineralized recalled | Medtronic Sofamor Danek USA | Other Bacterial Contamination | Nationwide |
More recalls from Medtronic Sofamor Danek USA
All 79| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Grafton" DBM and Grafton Plus" DBM contain demineralized recalled | Medtronic Sofamor Danek USA | Other Bacterial Contamination | Nationwide |
| Device | CLASS II | Catalyft PL & PL40 Expandable Interbody System labeled as recalled | Medtronic Sofamor Danek USA | Other | Nationwide |
| Device | CLASS II | Grafton DBM Crunch recalled by Medtronic Sofamor Danek USA | Medtronic Sofamor Danek USA | Processing Deviation | TN |
| Device | CLASS II | Grafton DBM Matrix Strips recalled by Medtronic Sofamor Danek USA | Medtronic Sofamor Danek USA | Processing Deviation | TN |
| Device | CLASS II | Grafton DBM Putty recalled by Medtronic Sofamor Danek USA | Medtronic Sofamor Danek USA | Processing Deviation | TN |
Other Hip & Knee Implants recalls
All 3,115| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | TS Femoral Prep Tray, Triathlon recalled by Howmedica Osteonics | Howmedica Osteonics | Sterility | NJ |
| Device | CLASS II | Grafton" DBM and Grafton Plus" DBM contain demineralized recalled | Medtronic Sofamor Danek USA | Other Bacterial Contamination | Nationwide |
| Device | CLASS II | Triathlon Hinge Femoral Prep Tray recalled by Howmedica Osteonics | Howmedica Osteonics | Sterility | NJ |
| Device | CLASS II | Intellijoint Navigation System: Workstation G2 Configured recalled | Intellijoint Surgical | Device Malfunction | Nationwide |
| Device | CLASS II | Trinity Non-Surgical Isolation Gown, AAMI PB70 Level 2, X-Large recalled | New York Embroidery Studio | Packaging Defects | Nationwide |