CLASS III Device recall · Reported 19 Dec 2018 · FDA Enforcement Report

Hemoglobin A1c Control recalled by Tosoh Bioscience

Tosoh Bioscience is recalling Hemoglobin A1c Control, distributed nationwide in the US. The recall was initiated on 5 Oct 2018 and the FDA has classified it as Class III.

Check your package

Lot, batch or date codesLot 7055
Quantity recalled1000

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0601-2019
FDA event ID
81188
Recalling firm
Tosoh Bioscience, Grove City, OH
Classification
Class III
Status
Terminated
Recall initiated
5 Oct 2018
FDA classified
13 Dec 2018
Reported
19 Dec 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay
Device type
Patient Monitors

Timeline

5 Oct 2018Recall initiated by company
13 Dec 2018Classified by FDA+69 days
19 Dec 2018Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Tosoh Bioscience has become aware of potential degradation of HbA1c control lot # 7055 before the 60-days product stability claim. If degradation of these controls occurs, P00 peaks may be seen and the % HbA1c control values may drift higher than the assigned range. This may result in delays when performing routine instrument quality control. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received two (2) complaints related to this issue with no serious injuries reported.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Hemoglobin A1c Control, Part no. 220232 for use as quality control material to monitor the precision of laboratory testing procedures for HbA1c quantitation.

Where it was distributed

Distributed nationwide. Foreign distribution to Canada.

Nationwide

Questions about this recall

Why was it recalled?

Tosoh Bioscience has become aware of potential degradation of HbA1c control lot # 7055 before the 60-days product stability claim. If degradation of these controls occurs, P00 peaks may be seen and the % HbA1c control values may drift higher than the assigned range. This may result in delays when performing routine instrument quality control. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received two (2) complaints related to this issue with no serious injuries reported.

Which lots are affected?

Lot 7055

Where was it sold?

Distributed nationwide. Foreign distribution to Canada.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class III mean?

Class III recalls are for products unlikely to cause any adverse health consequence, usually labelling or quality issues.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.