CLASS II Device recall · Reported 5 Feb 2020 · FDA Enforcement Report

High Speed Cell Saver¿5/5+ Bowl Set Autotransfusion Device recalled

Haemonetics is recalling High Speed Cell Saver¿5/5+ Bowl Set Autotransfusion Device, distributed nationwide in the US. The recall was initiated on 21 Oct 2019 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog number/FG Item Number: 0263A-00 1. No lots for cracking issue identified (0 Distributed). 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019.
Quantity recalled3563

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0887-2020
FDA event ID
84183
Recalling firm
Haemonetics, Braintree, MA
Classification
Class II
Status
Terminated
Recall initiated
21 Oct 2019
FDA classified
29 Jan 2020
Reported
5 Feb 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

21 Oct 2019Recall initiated by company
29 Jan 2020Classified by FDA+100 days
5 Feb 2020Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

High Speed Cell Saver¿5/5+ Bowl Set Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.

Where it was distributed

Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts

Questions about this recall

Why was it recalled?

Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

Which lots are affected?

Catalog number/FG Item Number: 0263A-00 1. No lots for cracking issue identified (0 Distributed). 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019.

Where was it sold?

Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

21 other products · FDA event 84183

This product is part of a recall event; the main page for the event is Japan Cell Saver Elite Set recalled by Haemonetics.

ReportedClassRecallCompanyReasonWhere
CLASS IIJapan Cell Saver Elite Set recalled by HaemoneticsHaemoneticsPackaging DefectsNationwide
CLASS IICS5/5+ Fastpacks, 20¿ Res Autotransfusion Device recalledHaemoneticsPackaging DefectsNationwide
CLASS IIJapan Disp Set, 261j, 205j, Har recalled by HaemoneticsHaemoneticsPackaging DefectsNationwide
CLASS IICS Elite Fastpack, 150 Res Autotransfusion Device recalledHaemoneticsPackaging DefectsNationwide
CLASS IIJapan DISP SET, 263J, 205J, HAR recalled by HaemoneticsHaemoneticsPackaging DefectsNationwide
CLASS IICS5/5+ Fastpacks, 20¿ Res Autotransfusion Device recalledHaemoneticsPackaging DefectsNationwide
CLASS IICS Elite Fastpack, 150 Res Autotransfusion Device recalledHaemoneticsPackaging DefectsNationwide
CLASS IICell Saver¿5/5+ Bowl Kit Autotransfusion Device recalled by HaemoneticsHaemoneticsPackaging DefectsNationwide
CLASS IIJapan Disp Set, 261j, 205j recalled by HaemoneticsHaemoneticsPackaging DefectsNationwide
CLASS IIJapan CS Elite Fastpack Disp recalled by HaemoneticsHaemoneticsPackaging DefectsNationwide
CLASS IICS Elite Processing Set Autotransfusion Device recalled by HaemoneticsHaemoneticsPackaging DefectsNationwide
CLASS IICS Elite Fastpack, 20¿ Res Autotransfusion Device recalledHaemoneticsPackaging DefectsNationwide
CLASS IICS5/5+ Bowl Kit- Autotransfusion Device recalled by HaemoneticsHaemoneticsPackaging DefectsNationwide
CLASS IICS5/5+ Fastpacks, 150¿ Res Autotransfusion Device recalledHaemoneticsPackaging DefectsNationwide
CLASS IIJapan Cs Elite Fastpack recalled by HaemoneticsHaemoneticsPackaging DefectsNationwide
CLASS IIJapan DISP SET, 263J, 205J recalled by HaemoneticsHaemoneticsPackaging DefectsNationwide
CLASS IICS5/5+ Fastpacks, 150¿ Res Autotransfusion Device recalledHaemoneticsPackaging DefectsNationwide
CLASS IICS Elite Processing Set Autotransfusion Device recalled by HaemoneticsHaemoneticsPackaging DefectsNationwide
CLASS IIJapan Cell Saver Elite Set recalled by HaemoneticsHaemoneticsPackaging DefectsNationwide
CLASS IIJapan Basic Hi Speed Cs5 Set,ln 263j recalled by HaemoneticsHaemoneticsPackaging DefectsNationwide
CLASS IIJapan Low Vol Cs5 Set,ln 261j recalled by HaemoneticsHaemoneticsPackaging DefectsNationwide

More recalls from Haemonetics

All 31

Other Other Medical Devices recalls

All 13,778