CLASS II ONGOING Device recall · Reported 23 Nov 2022 · FDA Enforcement Report
Hill-Rom Progressa Bed System recalled by Baxter Healthcare
Baxter Healthcare is recalling Hill-Rom Progressa Bed System, distributed nationwide in the US. The recall was initiated on 18 Oct 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0231-2023
- FDA event ID
- 90990
- Recalling firm
- Baxter Healthcare, Deerfield, IL
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 18 Oct 2022
- FDA classified
- 14 Nov 2022
- Reported
- 23 Nov 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 18 Oct 2022 | Recall initiated by company | |
| 14 Nov 2022 | Classified by FDA | +27 days |
| 23 Nov 2022 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The compression links on impacted Progressa beds, listed in Appendix A, may bend when raising the head section. The labeling for these beds list a maximum patient weight of 500 lbs., however the bending may still occur with a patient weight of less than 500 lbs. depending on the patient 's position on the bed.
Product as listed by the FDA
Hill-Rom Progressa Bed System
Where it was distributed
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, HI, IN, LA, MA, MO, NC, NM, NV, NY, OK, OR, PA, SC. TN, TX, WA, and WI. (UPDATED: States of PA and SC and there was government distribution.) There was no military distribution. Foreign distribution was made to Canada, Qatar, Singapore, and UAE. UPDATE: Saudi Arabia now included.
Nationwide ArizonaCaliforniaColoradoFloridaHawaiiIndianaLouisianaMassachusettsMissouriNorth CarolinaNew MexicoNevadaNew YorkOklahomaOregonPennsylvaniaSouth CarolinaTennesseeTexasWashington
Questions about this recall
Is Hill-Rom Progressa Bed System recalled?
Yes. Baxter Healthcare recalled Hill-Rom Progressa Bed System on 18 Oct 2022. The recall is Class II and its status is Ongoing. Recall number Z-0231-2023.
Why was it recalled?
The compression links on impacted Progressa beds, listed in Appendix A, may bend when raising the head section. The labeling for these beds list a maximum patient weight of 500 lbs., however the bending may still occur with a patient weight of less than 500 lbs. depending on the patient 's position on the bed.
Which lots are affected?
UDI/DI 00887761000100, Serial Numbers: X157AW7312, X206AW8017, X154AW7292, X166AW7416, X207AW8041, X158AW7351, X157AW7326, X157AW7313, X206AW8018, X154AW7293, X166AW7418, X167AW7449, X158AW7352, X157AW7328, X174AW7575, X206AW8019, X154AW7296, X167AW7428, X167AW7450, X158AW7353, X157AW7329, X174AW7579, X206AW8021, X154AW7297, X167AW7429, X167AW7451, X158AW7354, X157AW7330, X181AW7722, X206AW8022, X154AW7298, X157AW7308, X166AW7419, X158AW7355, X157AW7332, X181AW7728, X206AW8023, X154AW7300, X157AW7309, X167AW7425, X206AW8010, X158AW7340, X182AW7736, X206AW8024, X168AW7456, X178AW7622, X167AW7427, X206AW8011, X158AW7341, X182AW7737, X206AW8025, X168AW7478, X178AW7623, X159AW7375, X206AW8012, X158AW7342, X207AW8060, X158AW7344, X171AW7495, X178AW7625, X159AW7378, X206AW8020, X158AW7343, X208AW8079, X206AW8026, X171AW7496, X178AW7626, X159AW7380, X158AW7333, X158AW7349, X208AW8082, X206AW8027, X171AW7497, X178AW7627, X216AW8193, X158AW7334, X158AW7359, X208AW8084, X206AW8028, X171AW7498, X178AW7628, X216AW8194, X158AW7336, X159AW7360, X157AW7305, X206AW8029, X172AW7521, X178AW7629, X168AW7468, X158AW7337, X159AW7361, X158AW7348, X206AW8030, X172AW7522, X154AW7288, X168AW7474, X158AW7338, X159AW7364, X164AW7390, X207AW8038, X172AW7531, X154AW7295, X171AW7481, X158AW7356, X159AW7367, X166AW7403, X207AW8039, X172AW7533, X159AW7370, X206AW8032, X158AW7358, X159AW7374, X167AW7433, X207AW8040, X172AW7534, X159AW7371, X207AW8036, X159AW7363, X159AW7383, X167AW7434, X180AW7693, X173AW7536, X174AW7565, X167AW7447, X154AW7302, X159AW7384, X167AW7435, X181AW7714, X173AW7537, X174AW7567, X172AW7528, X157AW7304, X159AW7385, X201AW7935, X181AW7720, X159AW7382, X199AW7872, X172AW7529, X157AW7316, X159AW7386, X201AW7940, X154AW7289, X166AW7405, X158AW7345, X173AW7539, X157AW7317, X159AW7387, X203AW8005, X154AW7294, X166AW7406, X158AW7346, X173AW7540, X157AW7318, X164AW7395, X203AW8006, X164AW7391, X166AW7407, X158AW7347, X173AW7542, X157AW7320, X166AW7396, X207AW8042, X164AW7392, X166AW7411, X154AW7301, X173AW7543, X157AW7321, X166AW7398, X207AW8043, X164AW7388, X166AW7412, X157AW7311, X173AW7545, X157AW7322, X166AW7399, X207AW8044, X164AW7393, X166AW7414, X159AW7365, X173AW7546, X157AW7324, X166AW7410, X207AW8045, X154AW7291, X166AW7415, X159AW7366, X158AW7350, X157AW7325, X154AW7290 ADDITIONAL SERIAL NUMBERS UPDATED 2/14/2023: X140AW7268, X140AW7269, X140AW7270, X140AW7271, X140AW7272, X140AW7253, X140AW7254, X140AW7255, X140AW7256, X140AW7257, X140AW7258, X140AW7259, X140AW7260, X140AW7262, X140AW7263, X139AW7226, X139AW7227, X139AW7228, X139AW7229, X139AW7230, X103AW6603, X103AW6604, X103AW6605, X103AW6606, X103AW6607, X103AW6608, X103AW6609, X103AW6610, X103AW6611, X103AW6612, X103AW6613, X139AW7232, X139AW7234, X139AW7236, X139AW7237, X139AW7238, X139AW7239, X139AW7240, X139AW7241, X139AW7242, X139AW7243, X140AW7245, X140AW7246, X140AW7249, X140AW7250, X140AW7251, X140AW7252, X140AW7273, X140AW7274, X140AW7275, X154AW7280, X139AW7231, X139AW7235, X139AW7244, X140AW7261, X140AW7264, X140AW7265, X140AW7266, X140AW7267
Where was it sold?
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, HI, IN, LA, MA, MO, NC, NM, NV, NY, OK, OR, PA, SC. TN, TX, WA, and WI. (UPDATED: States of PA and SC and there was government distribution.) There was no military distribution. Foreign distribution was made to Canada, Qatar, Singapore, and UAE. UPDATE: Saudi Arabia now included.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Baxter Healthcare
All 564| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Drug | CLASS I | Dextrose Injection, USP, 70 % recalled by Baxter Healthcare | Baxter Healthcare | Particulate Matter | Nationwide |
| Drug | CLASS I | 9% Sodium Chloride Injection USP, Viaflex recalled by Baxter HealthcareSodium Chloride | Baxter Healthcare | Particulate Matter | IL |
| Device | CLASS II | Baxter Novum IQ Syringe Pump, Loaner recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Baxter Novum IQ Syringe Pump recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Drug | CLASS II | Clindamycin Phosphate in 9% Sodium Chloride Injection recalledClindamycin Phosphate | Baxter Healthcare | Sterility | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |