CLASS II ONGOING Device recall · Reported 6 Nov 2024 · FDA Enforcement Report
a. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed recalled
Baxter Healthcare is recalling a. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, distributed nationwide in the US. The recall was initiated on 18 Sep 2024 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0228-2025
- FDA event ID
- 95402
- Recalling firm
- Baxter Healthcare, Deerfield, IL
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 18 Sep 2024
- FDA classified
- 25 Oct 2024
- Reported
- 6 Nov 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Surgical Instruments
Timeline
| 18 Sep 2024 | Recall initiated by company | |
| 25 Oct 2024 | Classified by FDA | +37 days |
| 6 Nov 2024 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
a. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222, P7900B100281, P7900B1SPL05, P7900B000011, P7900B100018, P7900B100028, P7900B100223, P7900B100287, P7900B1SPL06, P7900B000020, P7900B100019, P7900B100039, P7900B100224, P7900B100288, P7900B000022, P7900B100020, P7900B100090, P7900B100231, P7900B100292, P7900B100001, P7900B100021, P7900B100161, P7900B100237, P7900B100293, P7900B100005, P7900B100022, P7900B100203, P7900B100242, P7900B100294, P7900B100009, P7900B100023, P7900B100205, P7900B100264, and P7900B100296. b. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900A000001, P7900A000010, P7900A000019, P7900A000039, P7900A000005, P7900A000011, P7900A000022, and P7900A000063.
Where it was distributed
Worldwide distribution - US Nationwide. There was government and military distribution. The countries of Canada and Australia.
Questions about this recall
Why was it recalled?
The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.
Which lots are affected?
All serial numbers with software version 1.27.000. UDI-DI 00887761985162: Product codes P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222, P7900B100281, P7900B1SPL05, P7900B000011, P7900B100018, P7900B100028, P7900B100223, P7900B100287, P7900B1SPL06, P7900B000020, P7900B100019, P7900B100039, P7900B100224, P7900B100288, P7900B000022, P7900B100020, P7900B100090, P7900B100231, P7900B100292, P7900B100001, P7900B100021, P7900B100161, P7900B100237, P7900B100293, P7900B100005, P7900B100022, P7900B100203, P7900B100242, P7900B100294, P7900B100009, P7900B100023, P7900B100205, P7900B100264, and P7900B100296. UDI-DI 00887761999961: Product codes P7900A000001, P7900A000010, P7900A000019, P7900A000039, P7900A000005, P7900A000011, P7900A000022, and P7900A000063.
Where was it sold?
Worldwide distribution - US Nationwide. There was government and military distribution. The countries of Canada and Australia.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Baxter Healthcare
All 577| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | PRO+ 36" MRS Surface, mattress recalled by Baxter Healthcare | Baxter Healthcare | Packaging Defects | Nationwide |
| Device | CLASS II | Pro+ Mrs Surface For Versacare, mattress recalled by Baxter Healthcare | Baxter Healthcare | Packaging Defects | Nationwide |
| Device | CLASS II | Volara System Volara, HC, NA recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | PRO+ 40" MRS Surface, mattress recalled by Baxter Healthcare | Baxter Healthcare | Packaging Defects | Nationwide |
| Drug | CLASS I | Dextrose Injection, USP, 70 % recalled by Baxter Healthcare | Baxter Healthcare | Particulate Matter | Nationwide |
Other Surgical Instruments recalls
All 3,891| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Adaptable Beach Chair Beach Chair surgical patient positioner recalled | Quantum OPS | Device Malfunction | Nationwide |
| Device | CLASS II | Laparoscopic Prostatectomy kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Halyard Laparoscopy Comprehensive Pack II recalled | GRI Medical & Electronic Technology | Other | Nationwide |
| Device | CLASS II | Gyn Laparoscopy/Laparotomy kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Gyn Laparoscopy Pack kits recalled by Avid Medical | Avid Medical | Other | Nationwide |