CLASS II ONGOING Device recall · Reported 6 Nov 2024 · FDA Enforcement Report

a. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed recalled

Baxter Healthcare is recalling a. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, distributed nationwide in the US. The recall was initiated on 18 Sep 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll serial numbers with software version 1.27.000. UDI-DI 00887761985162: Product codes P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222, P7900B100281, P7900B1SPL05, P7900B000011, P7900B100018, P7900B100028, P7900B100223, P7900B100287, P7900B1SPL06, P7900B000020, P7900B100019, P7900B100039, P7900B100224, P7900B100288, P7900B000022, P7900B100020, P7900B100090, P7900B100231, P7900B100292, P7900B100001, P7900B100021, P7900B100161, P7900B100237, P7900B100293, P7900B100005, P7900B100022, P7900B100203, P7900B100242, P7900B100294, P7900B100009, P7900B100023, P7900B100205, P7900B100264, and P7900B100296. UDI-DI 00887761999961: Product codes P7900A000001, P7900A000010, P7900A000019, P7900A000039, P7900A000005, P7900A000011, P7900A000022, and P7900A000063.
Quantity recalled2,967 beds

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0228-2025
FDA event ID
95402
Recalling firm
Baxter Healthcare, Deerfield, IL
Classification
Class II
Status
Ongoing
Recall initiated
18 Sep 2024
FDA classified
25 Oct 2024
Reported
6 Nov 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

18 Sep 2024Recall initiated by company
25 Oct 2024Classified by FDA+37 days
6 Nov 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

a. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222, P7900B100281, P7900B1SPL05, P7900B000011, P7900B100018, P7900B100028, P7900B100223, P7900B100287, P7900B1SPL06, P7900B000020, P7900B100019, P7900B100039, P7900B100224, P7900B100288, P7900B000022, P7900B100020, P7900B100090, P7900B100231, P7900B100292, P7900B100001, P7900B100021, P7900B100161, P7900B100237, P7900B100293, P7900B100005, P7900B100022, P7900B100203, P7900B100242, P7900B100294, P7900B100009, P7900B100023, P7900B100205, P7900B100264, and P7900B100296. b. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900A000001, P7900A000010, P7900A000019, P7900A000039, P7900A000005, P7900A000011, P7900A000022, and P7900A000063.

Where it was distributed

Worldwide distribution - US Nationwide. There was government and military distribution. The countries of Canada and Australia.

Nationwide

Questions about this recall

Why was it recalled?

The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.

Which lots are affected?

All serial numbers with software version 1.27.000. UDI-DI 00887761985162: Product codes P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222, P7900B100281, P7900B1SPL05, P7900B000011, P7900B100018, P7900B100028, P7900B100223, P7900B100287, P7900B1SPL06, P7900B000020, P7900B100019, P7900B100039, P7900B100224, P7900B100288, P7900B000022, P7900B100020, P7900B100090, P7900B100231, P7900B100292, P7900B100001, P7900B100021, P7900B100161, P7900B100237, P7900B100293, P7900B100005, P7900B100022, P7900B100203, P7900B100242, P7900B100294, P7900B100009, P7900B100023, P7900B100205, P7900B100264, and P7900B100296. UDI-DI 00887761999961: Product codes P7900A000001, P7900A000010, P7900A000019, P7900A000039, P7900A000005, P7900A000011, P7900A000022, and P7900A000063.

Where was it sold?

Worldwide distribution - US Nationwide. There was government and military distribution. The countries of Canada and Australia.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Baxter Healthcare

All 577

Other Surgical Instruments recalls

All 3,891