CLASS II Device recall · Reported 16 Jul 2014 · FDA Enforcement Report

HLD System recalled by Cenorin

Cenorin is recalling HLD System, distributed nationwide in the US. The recall was initiated on 24 Jan 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 611001, 611002, 611004, 611101, 611102, 611103, 611201, 611202, 611203, 610902L, 611204.
Quantity recalled25 HLD systems

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1995-2014
FDA event ID
68580
Recalling firm
Cenorin, Kent, WA
Classification
Class II
Status
Terminated
Recall initiated
24 Jan 2013
FDA classified
6 Jul 2014
Reported
16 Jul 2014
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Temperature Abuse

Timeline

24 Jan 2013Recall initiated by company
6 Jul 2014Classified by FDA+528 days
16 Jul 2014Published in enforcement report+10 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.

Foods that need refrigeration and were stored or shipped too warm can grow harmful bacteria without any change in look or smell.

Product as listed by the FDA

HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.

Where it was distributed

Worldwide Distribution - US (nationwide): CA, DE, FL, MA, MI, NY, PA, TX, UT, WA, and WI and the countries of Uruguay and Taiwan.

Nationwide CaliforniaDelawareFloridaMassachusettsMichiganNew YorkPennsylvaniaTexasUtahWashingtonWisconsin

Questions about this recall

Is HLD System recalled?

Yes. Cenorin recalled HLD System on 24 Jan 2013. The recall is Class II and its status is Terminated. Recall number Z-1995-2014.

Why was it recalled?

The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.

Which lots are affected?

610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 611001, 611002, 611004, 611101, 611102, 611103, 611201, 611202, 611203, 610902L, 611204.

Where was it sold?

Worldwide Distribution - US (nationwide): CA, DE, FL, MA, MI, NY, PA, TX, UT, WA, and WI and the countries of Uruguay and Taiwan.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Cenorin

All 6
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIHLD Systems 600 Series Washer/Pasteurizer recalled by CenorinCenorinProcessing DeviationNationwide
DeviceCLASS IIThermaSure(TM), a Cenorin(TM) product recalledCenorinOtherNationwide
DeviceCLASS IIThermaSure(TM), a Cenorin(TM) product recalledCenorinOtherNationwide
DeviceCLASS IIThermaSure(TM), a Cenorin(TM) product recalledCenorinOtherNationwide
DeviceCLASS IIHLD Systems recalled by CenorinCenorinTemperature AbuseNationwide

Other Other Medical Devices recalls

All 13,778