CLASS II Device recall · Reported 16 Jul 2014 · FDA Enforcement Report
HLD System recalled by Cenorin
Cenorin is recalling HLD System, distributed nationwide in the US. The recall was initiated on 24 Jan 2013 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1995-2014
- FDA event ID
- 68580
- Recalling firm
- Cenorin, Kent, WA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 24 Jan 2013
- FDA classified
- 6 Jul 2014
- Reported
- 16 Jul 2014
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Temperature Abuse
- Device type
- Other Medical Devices
Timeline
| 24 Jan 2013 | Recall initiated by company | |
| 6 Jul 2014 | Classified by FDA | +528 days |
| 16 Jul 2014 | Published in enforcement report | +10 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.
Foods that need refrigeration and were stored or shipped too warm can grow harmful bacteria without any change in look or smell.
Product as listed by the FDA
HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.
Where it was distributed
Worldwide Distribution - US (nationwide): CA, DE, FL, MA, MI, NY, PA, TX, UT, WA, and WI and the countries of Uruguay and Taiwan.
Nationwide CaliforniaDelawareFloridaMassachusettsMichiganNew YorkPennsylvaniaTexasUtahWashingtonWisconsin
Questions about this recall
Is HLD System recalled?
Yes. Cenorin recalled HLD System on 24 Jan 2013. The recall is Class II and its status is Terminated. Recall number Z-1995-2014.
Why was it recalled?
The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.
Which lots are affected?
610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 611001, 611002, 611004, 611101, 611102, 611103, 611201, 611202, 611203, 610902L, 611204.
Where was it sold?
Worldwide Distribution - US (nationwide): CA, DE, FL, MA, MI, NY, PA, TX, UT, WA, and WI and the countries of Uruguay and Taiwan.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Cenorin
All 6| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | HLD Systems 600 Series Washer/Pasteurizer recalled by Cenorin | Cenorin | Processing Deviation | Nationwide |
| Device | CLASS II | ThermaSure(TM), a Cenorin(TM) product recalled | Cenorin | Other | Nationwide |
| Device | CLASS II | ThermaSure(TM), a Cenorin(TM) product recalled | Cenorin | Other | Nationwide |
| Device | CLASS II | ThermaSure(TM), a Cenorin(TM) product recalled | Cenorin | Other | Nationwide |
| Device | CLASS II | HLD Systems recalled by Cenorin | Cenorin | Temperature Abuse | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |