CLASS II ONGOING Device recall · Reported 18 Nov 2026 · FDA Enforcement Report

Horizon QDR X-Ray Bone Densitometer recalled by Hologic

Hologic is recalling Horizon QDR X-Ray Bone Densitometer, distributed nationwide in the US. The recall was initiated on 20 Oct 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes1. Model Horizon W; UDI:15420045505834; Serial Numbers: 308912M, 308911M, 308909M, 308908M, 308905M, 308899M, 308892M, 308890M, 308887M, 308925M, 308921M, 308918M, 308969M, 308968M, 308965M, 308964M, 308960M, 308956M, 308947M, 308945M, 308944M, 308942M, 308940M, 308939M, 308936M, 308934M, 308932M, 308928M, 309026M, 309017M, 309033M, 309029M, 309066M, 309045M, 309071M, 309048M, 309094M, 309109M, 309106M, 309104M, 309099M, 309146M, 309144M, 309136M, 309133M, 309126M, 309122M, 309117M, 309113M, 309184M, 309181M, 309177M, 309150M, 309704M, 309702M, 309700M, 309709M, 309707M, 309719M, 309715M, 309713M, 309741M, 309740M, 309738M, 309736M, 309735M, 309734M, 309733M, 309732M, 309731M, 309730M, 309729M, 309727M, 309726M, 309725M, 309724M, 309723M, 309722M, 309717M, 309711M, 308823M, 309743M, 309739M, 309751M, 309749M, 309747M, 309745M; 2. Model Horizon WI; UDI: 15420045505865; Serial Numbers: 308848M, 308823M, 308818M, 308866M, 308862M, 308858M, 308855M, 308906M, 308902M, 308889M, 308884M, 309003M, 309002M, 309001M, 308946M, 308943M, 308941M, 308938M, 308937M, 308933M, 309008M, 309007M, 309006M, 309005M, 309004M, 309027M, 309023M, 309041M, 309037M, 309035M, 309032M, 309030M, 309069M, 309064M, 309062M, 309043M, 309039M, 309082M, 309080M, 309078M, 309076M, 309072M, 309070M, 309084M, 309187M, 309180M, 309176M, 309197M, 309196M, 309194M, 309193M, 309191M, 309189M, 309244M, 309243M, 309241M, 309240M, 309236M, 309265M, 309263M, 309251M, 309282M, 309255M, 309303M, 309298M, 309296M, 309293M, 309315M, 309312M, 309309M, 309306M, 309329M, 309326M, 309320M, 309346M, 309343M, 309340M, 309338M, 309336M, 309335M, 309362M, 309357M, 309356M, 309354M, 309374M, 309370M, 309359M, 309385M, 309379M, 309675M, 309703M, 309701M, 309699M, 309708M, 309705M, 309718M, 309716M, 309728M, 309742M, 309737M, 309721M, 309744M, 309714M, 309712M, 309710M, 309752M, 309750M, 309748M, 309746M, 309754M, 309753M;
Quantity recalled198

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0423-2027
FDA event ID
100116
Recalling firm
Hologic, Marlborough, MA
Classification
Class II
Status
Ongoing
Recall initiated
20 Oct 2026
FDA classified
12 Nov 2026
Reported
18 Nov 2026
Reason category
Device Malfunction

Timeline

20 Oct 2026Recall initiated by company
12 Nov 2026Classified by FDA+23 days
18 Nov 2026Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A nonconformance was identified for systems manufactured from December 2024 to January 2025 that were shipped without the validated VGA cables. Some installations were completed using VGA cables that may not meet Hologic s applicable performance and safety standards for this system.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Horizon QDR X-Ray Bone Densitometer; Models: 1. Horizon W, 2. Horizon-WI;

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Dominican Republic, France, Germany, Greece, India, Iran, Islamic Republic of, Israel, Italy, Japan, Korea, Republic of, Lebanon, Malaysia, Mexico, Morocco, Nepal, New Zealand, Pakistan, Peru, Philippines, Portugal, Romania, Saudi Arabia, Singapore, South Africa, Spain, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom.

Nationwide

Questions about this recall

Why was it recalled?

A nonconformance was identified for systems manufactured from December 2024 to January 2025 that were shipped without the validated VGA cables. Some installations were completed using VGA cables that may not meet Hologic s applicable performance and safety standards for this system.

Which lots are affected?

1. Model Horizon W; UDI:15420045505834; Serial Numbers: 308912M, 308911M, 308909M, 308908M, 308905M, 308899M, 308892M, 308890M, 308887M, 308925M, 308921M, 308918M, 308969M, 308968M, 308965M, 308964M, 308960M, 308956M, 308947M, 308945M, 308944M, 308942M, 308940M, 308939M, 308936M, 308934M, 308932M, 308928M, 309026M, 309017M, 309033M, 309029M, 309066M, 309045M, 309071M, 309048M, 309094M, 309109M, 309106M, 309104M, 309099M, 309146M, 309144M, 309136M, 309133M, 309126M, 309122M, 309117M, 309113M, 309184M, 309181M, 309177M, 309150M, 309704M, 309702M, 309700M, 309709M, 309707M, 309719M, 309715M, 309713M, 309741M, 309740M, 309738M, 309736M, 309735M, 309734M, 309733M, 309732M, 309731M, 309730M, 309729M, 309727M, 309726M, 309725M, 309724M, 309723M, 309722M, 309717M, 309711M, 308823M, 309743M, 309739M, 309751M, 309749M, 309747M, 309745M; 2. Model Horizon WI; UDI: 15420045505865; Serial Numbers: 308848M, 308823M, 308818M, 308866M, 308862M, 308858M, 308855M, 308906M, 308902M, 308889M, 308884M, 309003M, 309002M, 309001M, 308946M, 308943M, 308941M, 308938M, 308937M, 308933M, 309008M, 309007M, 309006M, 309005M, 309004M, 309027M, 309023M, 309041M, 309037M, 309035M, 309032M, 309030M, 309069M, 309064M, 309062M, 309043M, 309039M, 309082M, 309080M, 309078M, 309076M, 309072M, 309070M, 309084M, 309187M, 309180M, 309176M, 309197M, 309196M, 309194M, 309193M, 309191M, 309189M, 309244M, 309243M, 309241M, 309240M, 309236M, 309265M, 309263M, 309251M, 309282M, 309255M, 309303M, 309298M, 309296M, 309293M, 309315M, 309312M, 309309M, 309306M, 309329M, 309326M, 309320M, 309346M, 309343M, 309340M, 309338M, 309336M, 309335M, 309362M, 309357M, 309356M, 309354M, 309374M, 309370M, 309359M, 309385M, 309379M, 309675M, 309703M, 309701M, 309699M, 309708M, 309705M, 309718M, 309716M, 309728M, 309742M, 309737M, 309721M, 309744M, 309714M, 309712M, 309710M, 309752M, 309750M, 309748M, 309746M, 309754M, 309753M;

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Dominican Republic, France, Germany, Greece, India, Iran, Islamic Republic of, Israel, Italy, Japan, Korea, Republic of, Lebanon, Malaysia, Mexico, Morocco, Nepal, New Zealand, Pakistan, Peru, Philippines, Portugal, Romania, Saudi Arabia, Singapore, South Africa, Spain, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 25 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.