CLASS II ONGOING Device recall · Reported 20 May 2026 · FDA Enforcement Report
3dimensions Mammography System recalled by Hologic
Hologic is recalling 3dimensions Mammography System, distributed nationwide in the US. The recall was initiated on 8 Apr 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2168-2026
- FDA event ID
- 98566
- Recalling firm
- Hologic, Marlborough, MA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 8 Apr 2026
- FDA classified
- 12 May 2026
- Reported
- 20 May 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Imaging Equipment
Timeline
| 8 Apr 2026 | Recall initiated by company | |
| 12 May 2026 | Classified by FDA | +34 days |
| 20 May 2026 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Complaints have been received of systems developing loose, missing, or broken internal bolts over time.
Product as listed by the FDA
3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS-INTL3D, 3DM-SYS-INTL3D-MOB, 3DM-SYS-INTL3D-NS, 3DM-SYS-STD, 3DM-SYS-STD-MOB, 3DM-SYS-STD-NS.
Where it was distributed
US Nationwide distribution.
Questions about this recall
Why was it recalled?
Complaints have been received of systems developing loose, missing, or broken internal bolts over time.
Which lots are affected?
"[Model No.; UDI]: [3DM-SYS-INTL2D;15420045510609], [3DM-SYS-INTL2D-MOB;15420045510593], [3DM-SYS-INTL2D-NS;15420045510586], [3DM-SYS-INTL3D;15420045510579], [3DM-SYS-INTL3D-MOB;15420045510562], [3DM-SYS-INTL3D-NS;15420045510555], [3DM-SYS-STD;15420045510524], [3DM-SYS-STD-MOB;15420045510531], [3DM-SYS-STD-NS;15420045510548]; All Serial No.
Where was it sold?
US Nationwide distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
1 other product · FDA event 98566This product is part of a recall event; the main page for the event is Selenia Dimensions Mammography Systems recalled by Hologic.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Selenia Dimensions Mammography Systems recalled by Hologic | Hologic | Other | Nationwide |
More recalls from Hologic
All 48| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Selenia Dimensions Mammography Systems recalled by Hologic | Hologic | Other | Nationwide |
| Device | CLASS II | Genius" Review Station Display (Barco Monitor) recalled by Hologic | Hologic | Device Malfunction | Nationwide |
| Device | CLASS II | Aptima HPV Assay recalled by Hologic | Hologic | Potency & Assay | Nationwide |
| Device | CLASS II | Brevera Breast Biopsy System Disposable Needle recalled by Hologic | Hologic | Particulate Matter | Nationwide |
| Device | CLASS II | ThinPrep¿ CytoLyt¿ Solution recalled by Hologic | Hologic | Other | Nationwide |
Other Imaging Equipment recalls
All 2,341| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips EPIQ Elite Diagnostic Ultrasound System recalled | Philips Ultrasound | Labeling Errors | Nationwide |
| Device | CLASS II | GmbH Spectralis with Flex Module recalled by Heidelberg Engineering | Heidelberg Engineering | Potency & Assay | Nationwide |
| Device | CLASS II | Core Services Hub V8, a set of backend services including recalled | GE Healthcare | Other | Nationwide |
| Device | CLASS II | Centricity PACS V2, V3, V4, V6, V7 recalled by GE Healthcare | GE Healthcare | Device Malfunction | Nationwide |
| Device | CLASS I | Haylard Univ Imaging Biopsy Trucustom kits recalled by Avid Medical | Avid Medical | Sterility | 4 states |