CLASS II ONGOING Device recall · Reported 23 Sep 2026 · FDA Enforcement Report

IV Sets with drip chambers recalled by ICU Medical

ICU Medical is recalling IV Sets with drip chambers, distributed nationwide in the US. The recall was initiated on 12 May 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll products manufactured on or before 7/4/2026. UDI-DI: 20887787005831
Quantity recalled86,113

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-3162-2026
FDA event ID
99058
Recalling firm
ICU Medical, San Clemente, CA
Classification
Class II
Status
Ongoing
Recall initiated
12 May 2026
FDA classified
11 Sep 2026
Reported
23 Sep 2026
Type
Voluntary: Firm initiated
Customers notified by
Press Release
Reason category
Particulate Matter

Timeline

12 May 2026Recall initiated by company
11 Sep 2026Classified by FDA+122 days
23 Sep 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.

Product as listed by the FDA

IV Sets with drip chambers

Where it was distributed

Worldwide - US Nationwide distribution including in the states and territories of ND, TX, CT, KY, CO, AR, NC, NJ, VA, MD, OK, AK, IA, IN, KS, OR, CA, OH, WA, GA, AL, SC, MA, HI, FL, NH, NE, MN, MS, SD, MT, NY, DC, ID, WI, PA, AZ, MO, NV, TN, NM, WV, WY, IL, UT, MI, RI, VT, LA, DE, ME, VI, GU, PR and the countries of PK, PH, MY, HK, IN, DE, CR.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts

Questions about this recall

Why was it recalled?

Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.

Which lots are affected?

All products manufactured on or before 7/4/2026. UDI-DI: 20887787005831

Where was it sold?

Worldwide - US Nationwide distribution including in the states and territories of ND, TX, CT, KY, CO, AR, NC, NJ, VA, MD, OK, AK, IA, IN, KS, OR, CA, OH, WA, GA, AL, SC, MA, HI, FL, NH, NE, MN, MS, SD, MT, NY, DC, ID, WI, PA, AZ, MO, NV, TN, NM, WV, WY, IL, UT, MI, RI, VT, LA, DE, ME, VI, GU, PR and the countries of PK, PH, MY, HK, IN, DE, CR.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 29 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

7 other products · FDA event 99058

This product is part of a recall event; the main page for the event is Oncology sets with drip chambers recalled by ICU Medical.

More recalls from ICU Medical

All 663

Other Syringes & Needles recalls

All 1,065