CLASS II Device recall · Reported 8 Jan 2020 · FDA Enforcement Report

i-STAT cTnI cartridge recalled by Abbott Point Of Care

Abbott Point Of Care is recalling i-STAT cTnI cartridge, distributed in 29 states. The recall was initiated on 16 Oct 2019 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesList Number: 06P23-25 UDI: (01)00054749001986
Quantity recalled3,465,875

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0761-2020
FDA event ID
84166
Recalling firm
Abbott Point Of Care, Princeton, NJ
Classification
Class II
Status
Terminated
Recall initiated
16 Oct 2019
FDA classified
31 Dec 2019
Reported
8 Jan 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

16 Oct 2019Recall initiated by company
31 Dec 2019Classified by FDA+76 days
8 Jan 2020Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

i-STAT¿ cTnI blue cartridges may generate a higher than expected number of quality check codes. The observed rate of quality check codes could be in the 3 to 5% range.

Product as listed by the FDA

i-STAT cTnI cartridge

Where it was distributed

US: AK AL AZ CA CT DC FL GA HI IL IN KS KY MD MI MS NC NE NH NM NV NY OH OK OR PA SC SD TN TX VA WA WV WY

AlaskaAlabamaArizonaCaliforniaConnecticutDistrict of ColumbiaFloridaGeorgiaIllinoisKansasKentuckyMarylandMichiganMississippiNorth CarolinaNebraskaNew HampshireNew MexicoNevadaNew York

Questions about this recall

Why was it recalled?

i-STAT¿ cTnI blue cartridges may generate a higher than expected number of quality check codes. The observed rate of quality check codes could be in the 3 to 5% range.

Which lots are affected?

List Number: 06P23-25 UDI: (01)00054749001986

Where was it sold?

US: AK AL AZ CA CT DC FL GA HI IL IN KS KY MD MI MS NC NE NH NM NV NY OH OK OR PA SC SD TN TX VA WA WV WY

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.