CLASS I ONGOING Device recall · Reported 1 Jul 2026 · FDA Enforcement Report
Impella CP Set with SmartAssist (10th Generation) recalled by Abiomed
Abiomed is recalling Impella CP Set with SmartAssist (10th Generation) containing Low Profile Introducer Kits, distributed nationwide in the US. The recall was initiated on 22 May 2026 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2602-2026
- FDA event ID
- 99002
- Recalling firm
- Abiomed, Danvers, MA
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 22 May 2026
- FDA classified
- 25 Jun 2026
- Reported
- 1 Jul 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Heart & Vascular Devices
Timeline
| 22 May 2026 | Recall initiated by company | |
| 25 Jun 2026 | Classified by FDA | +34 days |
| 1 Jul 2026 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Potential for thrombus formation during prolonged use of the introducer.
Product as listed by the FDA
Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits 13 and 25 cm Japan. Product Code: 1000542.
Where it was distributed
Worldwide - US Nationwide distribution in the states of FL, IL, MN, NJ, NY, VA and the country of Japan.
Nationwide FloridaIllinoisMinnesotaNew JerseyNew YorkVirginia
Questions about this recall
Why was it recalled?
Potential for thrombus formation during prolonged use of the introducer.
Which lots are affected?
Impella Set Product Code: 1000542; GTIN: 813502013467; Serial numbers: N/A. Batch Number: 2041530. Introducer Product Code: 1000542; Batch Number: 2041530.
Where was it sold?
Worldwide - US Nationwide distribution in the states of FL, IL, MN, NJ, NY, VA and the country of Japan.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
3 other products · FDA event 99002This product is part of a recall event; the main page for the event is Impella CP Set with SmartAssist (10th Generation) recalled by Abiomed.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS I | Impella CP Set with SmartAssist (10th Generation) recalled by Abiomed | Abiomed | Other | Nationwide | |
| CLASS I | Low Profile Introducer Kit for Impella CP. Impella Set recalled | Abiomed | Other | Nationwide | |
| CLASS I | Low Profile Introducer Kit for Impella CP. Impella Set recalled | Abiomed | Other | Nationwide |
More recalls from Abiomed
All 74| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Automated Impella Controller (AIC) recalled by Abiomed | Abiomed | Potency & Assay | Nationwide |
| Device | CLASS I | Impella Optical Controller, Automated Impella Controller recalled | Abiomed | Device Malfunction | Nationwide |
| Device | CLASS I | Impella Optical Controller with the below product recalled by Abiomed | Abiomed | Device Malfunction | Nationwide |
| Device | CLASS I | Automated Impella Controller (AIC) recalled by Abiomed | Abiomed | Device Malfunction | Nationwide |
| Device | CLASS I | Introducer Kit, Pump Set recalled by Abiomed | Abiomed | Packaging Defects | Nationwide |
Other Heart & Vascular Devices recalls
All 6,220| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | TufTex¿ Over-the-Wire Embolectomy Catheter recalled | LeMaitre Vascular | Other | Nationwide |
| Device | CLASS II | Endovascular Pack kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Bard Ep Xt Decapolar Steerable, Diagnostic Ep Catheter, 6f recalled | Stryker Sustainability Solutions | Particulate Matter | Nationwide |
| Device | CLASS II | Medical convenience kits containing BD CliniClean Pericare recalled | American Contract Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Medical convenience kits containing BD CliniClean Pericare recalled | American Contract Systems | Device Malfunction | Nationwide |