CLASS II Device recall · Reported 6 Dec 2017 · FDA Enforcement Report

ImplantDirect Legacy(TM) 3 Implant L recalled

Implant Direct Sybron Manufacturing is recalling ImplantDirect Legacy(TM) 3 Implant L, distributed nationwide in the US. The recall was initiated on 10 Oct 2017 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Number: 79607
Quantity recalled75 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0185-2018
FDA event ID
78350
Recalling firm
Implant Direct Sybron Manufacturing, Westlake Village, CA
Classification
Class II
Status
Terminated
Recall initiated
10 Oct 2017
FDA classified
28 Nov 2017
Reported
6 Dec 2017
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Sterility

Timeline

10 Oct 2017Recall initiated by company
28 Nov 2017Classified by FDA+49 days
6 Dec 2017Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

One reported complaint and discovered that the outer vial was out of round and undersized. Therefore a portion of the opening may not allow inner vial and components to slide freely out of the vial. As a result, the cap could possibly not be screwed on properly and the sterilization compromised.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

ImplantDirect Legacy(TM) 3 Implant, 13MM L, REF 864713 This Legacy implant is a two-piece implant for one-stage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework.

Where it was distributed

Worldwide distribution: US (nationwide) including Puerto Rico and country of Japan.

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

One reported complaint and discovered that the outer vial was out of round and undersized. Therefore a portion of the opening may not allow inner vial and components to slide freely out of the vial. As a result, the cap could possibly not be screwed on properly and the sterilization compromised.

Which lots are affected?

Lot Number: 79607

Where was it sold?

Worldwide distribution: US (nationwide) including Puerto Rico and country of Japan.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Implant Direct Sybron Manufacturing

All 33

Other Surgical Instruments recalls

All 3,891