CLASS II Device recall · Reported 16 May 2018 · FDA Enforcement Report

Imris Imri 1.5T And 3T S System recalled by Deerfield Imaging

Deerfield Imaging is recalling Imris Imri 1.5T And 3T S System, distributed nationwide in the US. The recall was initiated on 4 Apr 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Numbers: 10111-00, 10073-01, 10043-01, 10097-01, 10006-11, 10076-21, 10081-01, 10089-01, 10101-01, 10093-01
Quantity recalled10 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1645-2018
FDA event ID
79816
Recalling firm
Deerfield Imaging, Minnetonka, MN
Classification
Class II
Status
Terminated
Recall initiated
4 Apr 2018
FDA classified
4 May 2018
Reported
16 May 2018
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Device Malfunction

Timeline

4 Apr 2018Recall initiated by company
4 May 2018Classified by FDA+30 days
16 May 2018Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The DC power to the IFOSS collision detector control panel at your facility is being adversely affected by the magnetic field. Our investigation shows the magnetic field is causing the IFOSS control panel to malfunction.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

IMRIS IMRI 1.5T AND 3T S System Product Usage: The IMRIS intraoperative MRI systems are traditional MRI systems that have been suspended on an overhead rail system. and is designated to operative inside an RF shielded room to facilitate intraoperative and multi room use.

Where it was distributed

Worldwide Distribution - US Nationwide in the states of OH, FL, NH, TX, Sweden, Canada

Nationwide FloridaNew HampshireTexas

Questions about this recall

Why was it recalled?

The DC power to the IFOSS collision detector control panel at your facility is being adversely affected by the magnetic field. Our investigation shows the magnetic field is causing the IFOSS control panel to malfunction.

Which lots are affected?

Serial Numbers: 10111-00, 10073-01, 10043-01, 10097-01, 10006-11, 10076-21, 10081-01, 10089-01, 10101-01, 10093-01

Where was it sold?

Worldwide Distribution - US Nationwide in the states of OH, FL, NH, TX, Sweden, Canada

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.