CLASS II Device recall · Reported 15 Apr 2015 · FDA Enforcement Report

Infx-8000V Interventional Angiography System recalled

Toshiba American Medical Systems is recalling Infx-8000V Interventional Angiography System, distributed nationwide in the US. The recall was initiated on 11 Mar 2015 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial No. U7E1462009, ACA1472001, UEB1352004, UFD1442017, AAA1492001, ACA1492002, ULD1412014, AHA1482001, UPB1372006, ULD1392013, U6A1472001, ULD1452016, UPB1352005, ULD1452015, UDE1352017, A4B1472002, U4A14X2001, UPB1342004, 99B1342133, 99B1342134, UFD1452018, UFD13X2015, UFD13X2016, UFD1362014, UUC1452012, A6A1492002, UUC13Z2011, AAA14X2002, A4A13Z2001.
Quantity recalled28 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1394-2015
FDA event ID
70748
Recalling firm
Toshiba American Medical Systems, Tustin, CA
Classification
Class II
Status
Terminated
Recall initiated
11 Mar 2015
FDA classified
6 Apr 2015
Reported
15 Apr 2015
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

11 Mar 2015Recall initiated by company
6 Apr 2015Classified by FDA+26 days
15 Apr 2015Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The table base rotational brake that holds the table top may not always be engaged when any force or side impact is applied to the table top.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

INFX-8000V Interventional Angiography System. CAT-880B catheterization table. Cardiac and Vascular intervention with fluoroscopic and fluorographic images. For the USA and Canada the system is registered as lnfinix CF-i or lnfinix VF-i, for national regulations. INFX-8000V series includes a floor mounted multi-axis positioner.

Where it was distributed

Nationwide Distribution

Nationwide

Questions about this recall

Why was it recalled?

The table base rotational brake that holds the table top may not always be engaged when any force or side impact is applied to the table top.

Which lots are affected?

Serial No. U7E1462009, ACA1472001, UEB1352004, UFD1442017, AAA1492001, ACA1492002, ULD1412014, AHA1482001, UPB1372006, ULD1392013, U6A1472001, ULD1452016, UPB1352005, ULD1452015, UDE1352017, A4B1472002, U4A14X2001, UPB1342004, 99B1342133, 99B1342134, UFD1452018, UFD13X2015, UFD13X2016, UFD1362014, UUC1452012, A6A1492002, UUC13Z2011, AAA14X2002, A4A13Z2001.

Where was it sold?

Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Toshiba American Medical Systems

All 57

Other Heart & Vascular Devices recalls

All 6,199