CLASS II Device recall · Reported 4 Jul 2018 · FDA Enforcement Report

Diagnostic Ultrasound System, Aplio I-series recalled

Toshiba American Medical Systems is recalling Diagnostic Ultrasound System, Aplio I-series, distributed nationwide in the US. The recall was initiated on 4 Jan 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel No. TUS-AI700. TUS-AI800, TUS-AI900. Serial No. HNA1792011, HAB1732009, HNA1772007, HNA1772006, HAB1732023, HAB1732013, HAB1732016, HAB16Z2002, NJC17X2086, NJC17X2087, JHA16X2004, HAB1732027, HAB16Z2001, HAB16Z2002, NJB1762003, JHA1752022, HAB16Z2003, HAB16Z2004, JHA1752019, NJB1782050, HAB1742033, NJB1772031, HNA1792013, HNA1772008, JHA1752010, HAB1742035, HAB1742036, NJB1762001, NJB1782042, JHA1762034, HAB1732030, HAB16Z2005, HAB1732007, HAB1742034, HAB1742037, JHA1752024, NJB1772019, JHA1762029, JHA1762031, NJB1782061, HAB1732019, HNA1772001, HAB1712008, HAB1722011, HAB1712010, HAB1712009, HNA1792014, HAB1722012, HAB16Z2003, HAB1732017, HAB1732018, HAB1712006, HAB1732014, HAB1712007, HAB1732015, HAB1732021, HAB1732020, HAB1732025, JHA1762030, HAB1732029, HAB1732031, NJC17X2066, HNA1772002, NJB1782060, NJB1782043, NJB1762002, JHA16X2002, NJB1782057, NJB1782055, JHA1752021, NJB1782058, NJB1762006, NJB1782056, NJC17X2089, NJB1772023, NJB1782051, NJB1782052, NJB1782053, NJB1782054, HAB1732006, NJB1782062, NJB1782063, NJB1762004, NJB1762005, NJB1772022, JHA1752014, JHA1752016, JHA1752017, JHA1752018, HNA1792012, NJC17X2068, NJC17X2069, NJC17X2070, NJC17X2067, NJB1782046, NJB1782047, NJB1782048, NJB1782049, NJB1772039, NJB1782041, NJB1782045, NJC17X2083, NJC17X2084, NJC17X2081, NJC17X2079, NJC17X2080, NJC17X2082, NJC17X2076, NJC17X2077, NJC17X2078, NJB1772021, NJB1782044, JHA1752020, HNA1772005, NJB1782064, JHA1762032, HAB1732010, HAB1732008, NJB1782059, NJB1762008, NJB1762007, NJB1762009, NJB1772024, NJB1772014, NJB1772017, NJB1772018, NJB1772015, NJB1772016, NJB1772011, NJB1772012, NJB1772020, NJB1762010, NJB1772035, NJB1772036, NJB1772037, NJB1772038, NJB1772026, NJB1772027, NJB1772028, NJB1772030, NJB1772034, NJB1772025, NJB1772032, NJB1772013, JHA1762036, NJC17X2090, JHA1762033, HNA1772010, HNA1772009, NJB1772029, HNA1772004, HNA1772003, NJB1782065, NJC17X2071, NJB17822040, NJC17X2074, NJC17X2075, NJC17X2085, HAB1732026, JHA16X2007, HAB1732022, JHA1752023, JHA1762026, JHA1762025, HAB1732028, JHA16X2001, JHA16X2003, HAB1742011, HAB1742012, JHA16X2001, JHA16X2003, JHA1752004, JHA1752005, JHA1752006, JHA1752007, JHA1752008, JHA1752009, JHA1752010, JHA1752011, JHA1752012, JHA1752013, NJB1772033, NJC17X2072, NJC17X2073, NJC17X2088, HAB16Z2001, HAB16Z2004, HAB16Z2005, HAB1732024, HAB1732032, HNA1792015, HNB17X2016, HNB17X2017, HNB17X2018, HNB17X2019, HNB17X2020, HNB17X2021, HNB17X2022, HNB17X2023, HNB17X2024, HNB17X2025, HNB17X2026, HNB17X2027, HNB17X2028, HNB17X2029, HNB17X2030, JHA16X2002, JHA16X2005, JHA16X2006, JHA16X2008, JHA16X2009, JHA1752011, JHA1752012, JHA1752013, JHA1752015, JHA1762027, JHA1762028, JHA1762035, JHA1762037, JHA1762038, JHA1762039, HAB1742001, HAB1742002, JHA16X2001
Quantity recalled162 units total

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2315-2018
FDA event ID
80019
Recalling firm
Toshiba American Medical Systems, Tustin, CA
Classification
Class II
Status
Terminated
Recall initiated
4 Jan 2018
FDA classified
28 Jun 2018
Reported
4 Jul 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

4 Jan 2018Recall initiated by company
28 Jun 2018Classified by FDA+175 days
4 Jul 2018Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Diagnostic Ultrasound System, Aplio I-series. Labeled as i700, i800, i900 The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.

Where it was distributed

USA (nationwide) Distribution to the states of : CA, CO, DE, GA, HI, IA, ID, IL, IN, MD, MI, MN, MO, NC, ND, NV, OH, OR, PA, SC, SD, TN, TX, VA, WA, WV and WY.

Nationwide CaliforniaColoradoDelawareGeorgiaHawaiiIowaIdahoIllinoisIndianaMarylandMichiganMinnesotaMissouriNorth CarolinaNorth DakotaNevadaOhioOregonPennsylvaniaSouth Carolina

Questions about this recall

Is Diagnostic Ultrasound System, Aplio I-series recalled?

Yes. Toshiba American Medical Systems recalled Diagnostic Ultrasound System, Aplio I-series on 4 Jan 2018. The recall is Class II and its status is Terminated. Recall number Z-2315-2018.

Why was it recalled?

The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.

Which lots are affected?

Model No. TUS-AI700. TUS-AI800, TUS-AI900. Serial No. HNA1792011, HAB1732009, HNA1772007, HNA1772006, HAB1732023, HAB1732013, HAB1732016, HAB16Z2002, NJC17X2086, NJC17X2087, JHA16X2004, HAB1732027, HAB16Z2001, HAB16Z2002, NJB1762003, JHA1752022, HAB16Z2003, HAB16Z2004, JHA1752019, NJB1782050, HAB1742033, NJB1772031, HNA1792013, HNA1772008, JHA1752010, HAB1742035, HAB1742036, NJB1762001, NJB1782042, JHA1762034, HAB1732030, HAB16Z2005, HAB1732007, HAB1742034, HAB1742037, JHA1752024, NJB1772019, JHA1762029, JHA1762031, NJB1782061, HAB1732019, HNA1772001, HAB1712008, HAB1722011, HAB1712010, HAB1712009, HNA1792014, HAB1722012, HAB16Z2003, HAB1732017, HAB1732018, HAB1712006, HAB1732014, HAB1712007, HAB1732015, HAB1732021, HAB1732020, HAB1732025, JHA1762030, HAB1732029, HAB1732031, NJC17X2066, HNA1772002, NJB1782060, NJB1782043, NJB1762002, JHA16X2002, NJB1782057, NJB1782055, JHA1752021, NJB1782058, NJB1762006, NJB1782056, NJC17X2089, NJB1772023, NJB1782051, NJB1782052, NJB1782053, NJB1782054, HAB1732006, NJB1782062, NJB1782063, NJB1762004, NJB1762005, NJB1772022, JHA1752014, JHA1752016, JHA1752017, JHA1752018, HNA1792012, NJC17X2068, NJC17X2069, NJC17X2070, NJC17X2067, NJB1782046, NJB1782047, NJB1782048, NJB1782049, NJB1772039, NJB1782041, NJB1782045, NJC17X2083, NJC17X2084, NJC17X2081, NJC17X2079, NJC17X2080, NJC17X2082, NJC17X2076, NJC17X2077, NJC17X2078, NJB1772021, NJB1782044, JHA1752020, HNA1772005, NJB1782064, JHA1762032, HAB1732010, HAB1732008, NJB1782059, NJB1762008, NJB1762007, NJB1762009, NJB1772024, NJB1772014, NJB1772017, NJB1772018, NJB1772015, NJB1772016, NJB1772011, NJB1772012, NJB1772020, NJB1762010, NJB1772035, NJB1772036, NJB1772037, NJB1772038, NJB1772026, NJB1772027, NJB1772028, NJB1772030, NJB1772034, NJB1772025, NJB1772032, NJB1772013, JHA1762036, NJC17X2090, JHA1762033, HNA1772010, HNA1772009, NJB1772029, HNA1772004, HNA1772003, NJB1782065, NJC17X2071, NJB17822040, NJC17X2074, NJC17X2075, NJC17X2085, HAB1732026, JHA16X2007, HAB1732022, JHA1752023, JHA1762026, JHA1762025, HAB1732028, JHA16X2001, JHA16X2003, HAB1742011, HAB1742012, JHA16X2001, JHA16X2003, JHA1752004, JHA1752005, JHA1752006, JHA1752007, JHA1752008, JHA1752009, JHA1752010, JHA1752011, JHA1752012, JHA1752013, NJB1772033, NJC17X2072, NJC17X2073, NJC17X2088, HAB16Z2001, HAB16Z2004, HAB16Z2005, HAB1732024, HAB1732032, HNA1792015, HNB17X2016, HNB17X2017, HNB17X2018, HNB17X2019, HNB17X2020, HNB17X2021, HNB17X2022, HNB17X2023, HNB17X2024, HNB17X2025, HNB17X2026, HNB17X2027, HNB17X2028, HNB17X2029, HNB17X2030, JHA16X2002, JHA16X2005, JHA16X2006, JHA16X2008, JHA16X2009, JHA1752011, JHA1752012, JHA1752013, JHA1752015, JHA1762027, JHA1762028, JHA1762035, JHA1762037, JHA1762038, JHA1762039, HAB1742001, HAB1742002, JHA16X2001

Where was it sold?

USA (nationwide) Distribution to the states of : CA, CO, DE, GA, HI, IA, ID, IL, IN, MD, MI, MN, MO, NC, ND, NV, OH, OR, PA, SC, SD, TN, TX, VA, WA, WV and WY.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Toshiba American Medical Systems

All 57

Other Imaging Equipment recalls

All 2,336