CLASS II ONGOING Device recall · Reported 6 Aug 2025 · FDA Enforcement Report

Ingenia Ambition recalled by Philips North America

Philips North America is recalling Ingenia Ambition X, distributed nationwide in the US. The recall was initiated on 30 Jun 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Numbers (REF): (1) 781356, (2) 782109, (3) 782138; UDI-DI: (1) 00884838090040, (2) 00884838098350, (3) 00884838108622; Serial Numbers: (1) 47000, 47001, 47002, 47003, 47004, 47006, 47007, 47009, 47011, 47012, 47013, 47014, 47015, 47016, 47017, 47018, 47019, 47020, 47021, 47022, 47023, 47024, 47025, 47026, 47027, 47028, 47029, 47030, 47032, 47033, 47034, 47035, 47037, 47038, 47039, 47040, 47041, 47043, 47044, 47045, 47047, 47048, 47049, 47050, 47051, 47052, 47053, 47054, 47056, 47057, 47058, 47059, 47060, 47061, 47062, 47063, 47064, 47065, 47066, 47067, 47068, 47070, 47071, 47072, 47074, 47075, 47077, 47078, 47080, 47081, 47082, 47083, 47084, 47085, 47088, 47089, 47090, 47091, 47092, 47093, 47094, 47095, 47096, 47097, 47098, 47099, 47173, 47174, 47175, 47176, 47177, 47183, 47184, 47185, 47187, 47188, 47189, 47190, 47192, 47193, 47194, 47195, 47196, 47197, 47198, 47199, 47200, 47202, 47203, 47204, 47205, 47206, 47207, 47208, 47209, 47210, 47211, 47215, 47216, 47219, 47220, 47221, 47222, 47225, 47226, 47227, 47230, 47231, 47233, 47237, 47242, 47244, 47245, 47248, 47251, 47252, 47253, 47254, 47255, 47258, 47260, 47262, 47264, 47268, 47269, 47270, 47271, 47272, 47274, 47275, 47276, 47278, 47279, 47280, 47281, 47282, 47283, 47284, 47286, 47288, 47289, 47290, 47291, 47292, 47294, 47297, 47298, 47299, 47301, 47302, 47303, 47304, 47305, 47306, 47307, 47309, 47310, 47311, 47312, 47313, 47314, 47315, 47316, 47322, 47323, 47324, 47325, 47328, 47329, 47330, 47331, 47332, 47333, 47335, 47336, 47337, 47338, 47339, 47340, 47341, 47342, 47344, 47345, 47346, 47347, 47348, 47350, 47351, 47352, 47353, 47356, 47357, 47361, 47364, 47365, 47367, 47370, 47371, 47372, 47377, 47378, 47380, 47382, 47384, 47385, 47387, 47506, (2) 47500, 47502, 47503, 47504, 47505, 47507, 47508, 47509, 47511, 47512, 47513, 47514, 47516, 47517, 47519, 47520, 47521, 47522, 47523, 47525, 47526, 47527, 47528, 47529, 47530, 47531, 47532, 47533, 47534, 47535, 47536, 47538, 47539, 47540, 47541, 47542, 47543, 47545, 47546, 47547, 47548, 47549, 47550, 47552, 47553, 47554, 47555, 47557, 47558, 47559, 47560, 47561, 47562, 47563, 47564, 47565, 47566, 47567, 47568, 47569, 47570, 47572, 47574, 47575, 47576, 47577, 47578, 47579, 47580, 47582, 47583, 47584, 47585, 47588, 47589, 47590, 47591, 47594, 47595, 47596, 47597, 47598, 47599, 47602, 47603, 47606, 47607, 47608, 47609, 47610, 47612, 47614, 47615, 47617, 47621, 47622, 47623, 47625, 47626, 47627, 47628, 47629, 47630, 47632, 47634, 47635, 47640, 47642, 47644, 47646, 47648, 47649, 47651, 47652, 47653, 47654, 47656, 47657, 47659, 47660, 47661, 47662, 47663, 47665, 47666, 47667, 47668, 47669, 47671, 47672, 47674, 47675, 47676, 47677, 47678, 47679, 47680, 47681, 47682, 47683, 47684, 47685, 47687, 47688, 47689, 47692, 47693, 47694, 47695, 47696, 47697, 47698, 47699, 47700, 47701, 47702, 47703, 47705, 47706, 47708, 47709, 47710, 47711, 47713, 47714, 47716, 47717, 47718, 47719, 47720, 47721, 47722, 47723, 47724, 47725, 47726, 47727, 47728, 47729, 47730, 47731, 47732, 47733, 47737, 47738, 47739, 47740, 47742, 47747, 47748, 47750, 47751, 47752, 47753, 47754, 47755, 47759, 47760, 47761, 47762, 47763, 47765, 47766, 47767, 47768, 47770, 47771, 47773, 47778, 47779, 47780, 47781, 47782, 47783, 47784, 47785, 47786, 47788, 47789, 47790, 47791, 47792, 47794, 47795, 47796, 47797, 47798, 47799, 47800, 47801, 47802, 47804, 47806, 47807, 47808, 47809, 47810, 47811, 47812, 47813, 47814, 47815, 47816, 47817, 47818, 47819, 47820, 47822, 47823, 47824, 47826, 47831, 47834, 47835, 47839, 47841, 47843, 47844, 47845, 47846, 47847, 47848, 47849, 47850, 47851, 47853, 47854, 47855, 47858, 47859, 47860, 47861, 47862, 47863, 47864, 47865, 47866, 47867, 47868, 47869, 47870, 47874, 47876, 47877, 47878, 47879, 47881, 47882, 47883, 47884, 47885, 47886, 47887, 47888, 47889, 47891, 47892, 47893, 47894, 47896, 47899, 47900, 47901, 47902, 47903, 47904, 47905, 47906, 47907, 47909, 47910, 47912, 47914, 47919, 47921, 47923, 47928, 47929, 47931, 47932, 47936, 47937, (3) 49000, 49001, 49002, 49003, 49004, 49008, 49009, 49010, 49013, 49014, 49015, 49017, 49019, 49020, 49021, 49025, 49026, 49027, 49029, 49031, 49033, 49034, 49036, 49043, 49044, 49045, 49047, 49048, 49049, 49052, 49053, 49054, 49056, 49057, 49058, 49060, 49062, 49063, 49064, 49066, 49067, 49068, 49069, 49070, 49071, 49072, 49073, 49075, 49076, 49077, 49078, 49079, 49080, 49081, 49086, 49087, 49088, 49089, 49090, 49091, 49092, 49093, 49094, 49095, 49096, 49097, 49098, 49099, 49100, 49101, 49102, 49103, 49104, 49105, 49106, 49107, 49108, 49109, 49110, 49111, 49113, 49114, 49115, 49116, 49117, 49118, 49119, 49121, 49124, 49128, 49129, 49130, 49131, 49134, 49136, 49140, 49144, 49145, 49146, 49147, 49148, 49152, 49153, 49154, 49155, 49156, 49157, 49158, 49160, 49161, 49162, 49163, 49164, 49165, 49166, 49168, 49171, 49172, 49178, 49179, 49180, 49181, 49182, 49183, 49184, 49185, 49186, 49197, 49203, 49204, 49205, 49211, 49212, 49215, 49217, 49222, 49223, 49226, 49227, 49231, 49232;
Quantity recalled690 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2209-2025
FDA event ID
97172
Recalling firm
Philips North America, Cambridge, MA
Classification
Class II
Status
Ongoing
Recall initiated
30 Jun 2025
FDA classified
30 Jul 2025
Reported
6 Aug 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

30 Jun 2025Recall initiated by company
30 Jul 2025Classified by FDA+30 days
6 Aug 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Ingenia Ambition X Model Numbers (REF): (1) 781356, (2) 782109, (3) 782138;

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, French Guiana, FrenchPolynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Lesotho, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen, Zambia.

Nationwide Georgia

Questions about this recall

Why was it recalled?

The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.

Which lots are affected?

Model Numbers (REF): (1) 781356, (2) 782109, (3) 782138; UDI-DI: (1) 00884838090040, (2) 00884838098350, (3) 00884838108622; Serial Numbers: (1) 47000, 47001, 47002, 47003, 47004, 47006, 47007, 47009, 47011, 47012, 47013, 47014, 47015, 47016, 47017, 47018, 47019, 47020, 47021, 47022, 47023, 47024, 47025, 47026, 47027, 47028, 47029, 47030, 47032, 47033, 47034, 47035, 47037, 47038, 47039, 47040, 47041, 47043, 47044, 47045, 47047, 47048, 47049, 47050, 47051, 47052, 47053, 47054, 47056, 47057, 47058, 47059, 47060, 47061, 47062, 47063, 47064, 47065, 47066, 47067, 47068, 47070, 47071, 47072, 47074, 47075, 47077, 47078, 47080, 47081, 47082, 47083, 47084, 47085, 47088, 47089, 47090, 47091, 47092, 47093, 47094, 47095, 47096, 47097, 47098, 47099, 47173, 47174, 47175, 47176, 47177, 47183, 47184, 47185, 47187, 47188, 47189, 47190, 47192, 47193, 47194, 47195, 47196, 47197, 47198, 47199, 47200, 47202, 47203, 47204, 47205, 47206, 47207, 47208, 47209, 47210, 47211, 47215, 47216, 47219, 47220, 47221, 47222, 47225, 47226, 47227, 47230, 47231, 47233, 47237, 47242, 47244, 47245, 47248, 47251, 47252, 47253, 47254, 47255, 47258, 47260, 47262, 47264, 47268, 47269, 47270, 47271, 47272, 47274, 47275, 47276, 47278, 47279, 47280, 47281, 47282, 47283, 47284, 47286, 47288, 47289, 47290, 47291, 47292, 47294, 47297, 47298, 47299, 47301, 47302, 47303, 47304, 47305, 47306, 47307, 47309, 47310, 47311, 47312, 47313, 47314, 47315, 47316, 47322, 47323, 47324, 47325, 47328, 47329, 47330, 47331, 47332, 47333, 47335, 47336, 47337, 47338, 47339, 47340, 47341, 47342, 47344, 47345, 47346, 47347, 47348, 47350, 47351, 47352, 47353, 47356, 47357, 47361, 47364, 47365, 47367, 47370, 47371, 47372, 47377, 47378, 47380, 47382, 47384, 47385, 47387, 47506, (2) 47500, 47502, 47503, 47504, 47505, 47507, 47508, 47509, 47511, 47512, 47513, 47514, 47516, 47517, 47519, 47520, 47521, 47522, 47523, 47525, 47526, 47527, 47528, 47529, 47530, 47531, 47532, 47533, 47534, 47535, 47536, 47538, 47539, 47540, 47541, 47542, 47543, 47545, 47546, 47547, 47548, 47549, 47550, 47552, 47553, 47554, 47555, 47557, 47558, 47559, 47560, 47561, 47562, 47563, 47564, 47565, 47566, 47567, 47568, 47569, 47570, 47572, 47574, 47575, 47576, 47577, 47578, 47579, 47580, 47582, 47583, 47584, 47585, 47588, 47589, 47590, 47591, 47594, 47595, 47596, 47597, 47598, 47599, 47602, 47603, 47606, 47607, 47608, 47609, 47610, 47612, 47614, 47615, 47617, 47621, 47622, 47623, 47625, 47626, 47627, 47628, 47629, 47630, 47632, 47634, 47635, 47640, 47642, 47644, 47646, 47648, 47649, 47651, 47652, 47653, 47654, 47656, 47657, 47659, 47660, 47661, 47662, 47663, 47665, 47666, 47667, 47668, 47669, 47671, 47672, 47674, 47675, 47676, 47677, 47678, 47679, 47680, 47681, 47682, 47683, 47684, 47685, 47687, 47688, 47689, 47692, 47693, 47694, 47695, 47696, 47697, 47698, 47699, 47700, 47701, 47702, 47703, 47705, 47706, 47708, 47709, 47710, 47711, 47713, 47714, 47716, 47717, 47718, 47719, 47720, 47721, 47722, 47723, 47724, 47725, 47726, 47727, 47728, 47729, 47730, 47731, 47732, 47733, 47737, 47738, 47739, 47740, 47742, 47747, 47748, 47750, 47751, 47752, 47753, 47754, 47755, 47759, 47760, 47761, 47762, 47763, 47765, 47766, 47767, 47768, 47770, 47771, 47773, 47778, 47779, 47780, 47781, 47782, 47783, 47784, 47785, 47786, 47788, 47789, 47790, 47791, 47792, 47794, 47795, 47796, 47797, 47798, 47799, 47800, 47801, 47802, 47804, 47806, 47807, 47808, 47809, 47810, 47811, 47812, 47813, 47814, 47815, 47816, 47817, 47818, 47819, 47820, 47822, 47823, 47824, 47826, 47831, 47834, 47835, 47839, 47841, 47843, 47844, 47845, 47846, 47847, 47848, 47849, 47850, 47851, 47853, 47854, 47855, 47858, 47859, 47860, 47861, 47862, 47863, 47864, 47865, 47866, 47867, 47868, 47869, 47870, 47874, 47876, 47877, 47878, 47879, 47881, 47882, 47883, 47884, 47885, 47886, 47887, 47888, 47889, 47891, 47892, 47893, 47894, 47896, 47899, 47900, 47901, 47902, 47903, 47904, 47905, 47906, 47907, 47909, 47910, 47912, 47914, 47919, 47921, 47923, 47928, 47929, 47931, 47932, 47936, 47937, (3) 49000, 49001, 49002, 49003, 49004, 49008, 49009, 49010, 49013, 49014, 49015, 49017, 49019, 49020, 49021, 49025, 49026, 49027, 49029, 49031, 49033, 49034, 49036, 49043, 49044, 49045, 49047, 49048, 49049, 49052, 49053, 49054, 49056, 49057, 49058, 49060, 49062, 49063, 49064, 49066, 49067, 49068, 49069, 49070, 49071, 49072, 49073, 49075, 49076, 49077, 49078, 49079, 49080, 49081, 49086, 49087, 49088, 49089, 49090, 49091, 49092, 49093, 49094, 49095, 49096, 49097, 49098, 49099, 49100, 49101, 49102, 49103, 49104, 49105, 49106, 49107, 49108, 49109, 49110, 49111, 49113, 49114, 49115, 49116, 49117, 49118, 49119, 49121, 49124, 49128, 49129, 49130, 49131, 49134, 49136, 49140, 49144, 49145, 49146, 49147, 49148, 49152, 49153, 49154, 49155, 49156, 49157, 49158, 49160, 49161, 49162, 49163, 49164, 49165, 49166, 49168, 49171, 49172, 49178, 49179, 49180, 49181, 49182, 49183, 49184, 49185, 49186, 49197, 49203, 49204, 49205, 49211, 49212, 49215, 49217, 49222, 49223, 49226, 49227, 49231, 49232;

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, French Guiana, FrenchPolynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Lesotho, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen, Zambia.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

18 other products · FDA event 97172

This product is part of a recall event; the main page for the event is Ingenia Elition recalled by Philips North America.

ReportedClassRecallCompanyReasonWhere
CLASS IIIngenia Elition recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IISmartPath to dStream for 3.0T recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIngenia 1.5T recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIngenia 3.0T recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIMR 5300 Model Numbers (REF): (1) 782110, (2) recalledPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIngenia Ambition recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIAchieva 3.0T recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIUpgrade to MR recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIngenia 3.0T CX recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIngenia Elition recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIAchieva XR recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IISmartPath to Ingenia Elition recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIEvolution Upgrade 3.0T recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIEvolution Upgrade 1.5T recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntera 3.0T Quasar Dual recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIngenia 1.5T recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IISmartPath to dStream for XR and 3.0T recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIMR 7700 Model Numbers (REF): (1) 782120, (2) recalledPhilips North AmericaDevice MalfunctionNationwide

More recalls from Philips North America

All 531

Other Other Medical Devices recalls

All 13,778