CLASS II Device recall · Reported 24 Apr 2013 · FDA Enforcement Report
Innova 2100IQ, Cardiovascular and Interventional Imaging System recalled
GE Healthcare is recalling Innova 2100IQ, Cardiovascular and Interventional Imaging System, distributed nationwide in the US. The recall was initiated on 21 Oct 2011 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1142-2013
- FDA event ID
- 64884
- Recalling firm
- GE Healthcare, Waukesha, WI
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 21 Oct 2011
- FDA classified
- 18 Apr 2013
- Reported
- 24 Apr 2013
- Type
- Voluntary: Firm initiated
- Customers notified by
- Visit
- Reason category
- Other Reasons
- Device type
- Heart & Vascular Devices
Timeline
| 21 Oct 2011 | Recall initiated by company | |
| 18 Apr 2013 | Classified by FDA | +545 days |
| 24 Apr 2013 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
As part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In the situation a staff member (doctor, nurse, operator, anesthetist, etc&) stands in the region between the C-arm gantry pivot and the head section of the table, there is a risk of co
Product as listed by the FDA
GE Healthcare, Innova 2100IQ, Innova 3100IQ, Innova 4100IQ, Cardiovascular and Interventional Imaging System.
Where it was distributed
Worldwide Distribution - Nationwide Distribution including the states of FL, IL, IN, LA, MI, MN, NJ, NM, NY, OH, PR, SC, TX and VA and the countries of Vietnam, Latvia, Italy, Israel, Indonesia, Hong Kong, United Kingdom, Finland, Spain, Czech Republic, Canada, Bahrain and Australia.
Nationwide FloridaIllinoisIndianaLouisianaMichiganMinnesotaNew JerseyNew MexicoNew YorkOhioPuerto RicoSouth CarolinaTexasVirginia
Questions about this recall
Why was it recalled?
As part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In the situation a staff member (doctor, nurse, operator, anesthetist, etc&) stands in the region between the C-arm gantry pivot and the head section of the table, there is a risk of co
Which lots are affected?
Serial Number 00000000321WVO 00000000353WV3 00000000000000 00000000346WV7 00000000292WW3 00000000327WV7 00000000333WV5 00000000325WV1 00000000298WV0 00000000301WV2 00000000271WV7 00000000250WV1 00000000251WV9 00000000332WV7 00000000263WV4 00000000275WV8 00000000335WV0 00000000249WV3 00000000303WV8 00000000331WV9 00000000255WV0 00000000337WV6 00000000359WV0 00000000362WV4 00000000352WV5 00000000318WV6 00000000324WV4 00000000355WV8 00000000354WV1 00000000329WV3 00000000274WV1 00000000267WV5 00000000305WV3 00000000322WV8 00000000317WV8 00000000132WV1 00000000280WV8 00000000026WV5 00000000254WV3 00000000420WV0 00000000297WV2 00000XXX299WV8 00000000257WV6 00000000357WV4 00000000253WV5 00000000339WV2 00000000265WV9 00000000252WV7
Where was it sold?
Worldwide Distribution - Nationwide Distribution including the states of FL, IL, IN, LA, MI, MN, NJ, NM, NY, OH, PR, SC, TX and VA and the countries of Vietnam, Latvia, Italy, Israel, Indonesia, Hong Kong, United Kingdom, Finland, Spain, Czech Republic, Canada, Bahrain and Australia.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from GE Healthcare
All 512| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Signa Premier systems recalled by GE Healthcare | GE Healthcare | Other | Nationwide |
| Device | CLASS II | Centricity Universal Viewer Zero Footprint Client recalled | GE Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | GE HealthCare ViewPoint 6, Catalog Number H47601AA; Radiological recalled | GE Healthcare | Device Malfunction | |
| Device | CLASS II | Centricity Pacs-iw (Pacs-iw) recalled by GE Healthcare | GE Healthcare | Other | IL |
| Device | CLASS II | Centricity Pacs-iw with Universal Viewer recalled by GE Healthcare | GE Healthcare | Other | IL |
Other Heart & Vascular Devices recalls
All 6,199| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Getinge PVC Thoracic Catheter, Soft PVC Right Angle recalled | Atrium Medical | Potency & Assay | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes: *Bardex¿ I.C recalled | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Getinge PVC Thoracic Catheter, Firm PVC Straight recalled | Atrium Medical | Potency & Assay | Nationwide |