CLASS II ONGOING Device recall · Reported 10 Sep 2025 · FDA Enforcement Report
InPen App, Cfn: Mmt-8060 (iOS Users) and Mmt-8061 (Android Users) recalled
Medtronic MiniMed is recalling InPen App, Cfn: Mmt-8060 (iOS Users) and Mmt-8061 (Android Users), distributed in Georgia. The recall was initiated on 16 Jun 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2496-2025
- FDA event ID
- 97237
- Recalling firm
- Medtronic MiniMed, Northridge, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 16 Jun 2025
- FDA classified
- 2 Sep 2025
- Reported
- 10 Sep 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Medical Software
Timeline
| 16 Jun 2025 | Recall initiated by company | |
| 2 Sep 2025 | Classified by FDA | +78 days |
| 10 Sep 2025 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Medtronic MiniMed, Inc. is recalling InPen App for iOS and Android users due to software design errors that could lead to a missed short-acting insulin dose reminder and a recommendation to correct a high glucose value. It does not impact insulin delivery, long-acting insulin dose reminders, or CGM alerts, and users can still use the pen itself to calculate a dose, deliver insulin, record the dose date/time, and view CGM data. This issue was identified during internal testing before release in the US but after release to OUS customers, no complaints or MDRs related to this recall have been reported. Use of the affected device may result hyperglycemia by failing to alert the user and delay treatment of diabetes.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
InPen App, CFN: MMT-8060 (iOS Users) and MMT-8061 (Android Users)
Where it was distributed
International distribution to the countries of Argentina, Australia, Austria, Belgium, Bermuda, Chile, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Luxembourg, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
Questions about this recall
Why was it recalled?
Medtronic MiniMed, Inc. is recalling InPen App for iOS and Android users due to software design errors that could lead to a missed short-acting insulin dose reminder and a recommendation to correct a high glucose value. It does not impact insulin delivery, long-acting insulin dose reminders, or CGM alerts, and users can still use the pen itself to calculate a dose, deliver insulin, record the dose date/time, and view CGM data. This issue was identified during internal testing before release in the US but after release to OUS customers, no complaints or MDRs related to this recall have been reported. Use of the affected device may result hyperglycemia by failing to alert the user and delay treatment of diabetes.
Which lots are affected?
UDI-DI: 0763000B000122075 CFN/Software Version: MMT-8060/7.0.0, 7.1.0, 7.2.0 and 7.5.0 MMT-8061/7.0.0, 7.0.1, 7.1.0, 7.1.1, 7.2.0 and 7.5.0
Where was it sold?
International distribution to the countries of Argentina, Australia, Austria, Belgium, Bermuda, Chile, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Luxembourg, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Medtronic MiniMed
All 81| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | CareLink Clinic recalled by Medtronic MiniMed | Medtronic MiniMed | Device Malfunction | Nationwide |
| Device | CLASS II | The MiniMed Go app for iOS or MiniMed Go app recalled | Medtronic MiniMed | Device Malfunction | Nationwide |
| Device | CLASS II | MiniMed Flex Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Labeling Errors | Nationwide |
| Device | CLASS II | MiniMed Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Other | CA |
| Device | CLASS II | MiniMed Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Other | CA |
Other Medical Software recalls
All 935| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Metatarsophalangeal (MTP) Implant is a sterile recalled by restor3d | restor3d | Sterility | Nationwide |
| Device | CLASS II | Ion Torrent Genexus Dx recalled by Life Technologies Holdings Pte | Life Technologies Holdings Pte | Device Malfunction | Nationwide |
| Device | CLASS II | Mulitple-type osteoarticular pathogen nucleic acid IVD, kit recalled | BioFire Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS III | Kit: Urinary Clean Catch 20/cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Device Malfunction | CA |
| Device | CLASS II | Spectral CT and Spectral CT 7500 RT recalled by Philips Medical Systems | Philips Medical Systems | Device Malfunction | Nationwide |