CLASS II ONGOING Device recall · Reported 10 Sep 2025 · FDA Enforcement Report

InPen App, Cfn: Mmt-8060 (iOS Users) and Mmt-8061 (Android Users) recalled

Medtronic MiniMed is recalling InPen App, Cfn: Mmt-8060 (iOS Users) and Mmt-8061 (Android Users), distributed in Georgia. The recall was initiated on 16 Jun 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 0763000B000122075 CFN/Software Version: MMT-8060/7.0.0, 7.1.0, 7.2.0 and 7.5.0 MMT-8061/7.0.0, 7.0.1, 7.1.0, 7.1.1, 7.2.0 and 7.5.0
Quantity recalled6816

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2496-2025
FDA event ID
97237
Recalling firm
Medtronic MiniMed, Northridge, CA
Classification
Class II
Status
Ongoing
Recall initiated
16 Jun 2025
FDA classified
2 Sep 2025
Reported
10 Sep 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Medical Software

Timeline

16 Jun 2025Recall initiated by company
2 Sep 2025Classified by FDA+78 days
10 Sep 2025Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Medtronic MiniMed, Inc. is recalling InPen App for iOS and Android users due to software design errors that could lead to a missed short-acting insulin dose reminder and a recommendation to correct a high glucose value. It does not impact insulin delivery, long-acting insulin dose reminders, or CGM alerts, and users can still use the pen itself to calculate a dose, deliver insulin, record the dose date/time, and view CGM data. This issue was identified during internal testing before release in the US but after release to OUS customers, no complaints or MDRs related to this recall have been reported. Use of the affected device may result hyperglycemia by failing to alert the user and delay treatment of diabetes.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

InPen App, CFN: MMT-8060 (iOS Users) and MMT-8061 (Android Users)

Where it was distributed

International distribution to the countries of Argentina, Australia, Austria, Belgium, Bermuda, Chile, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Luxembourg, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.

Georgia

Questions about this recall

Why was it recalled?

Medtronic MiniMed, Inc. is recalling InPen App for iOS and Android users due to software design errors that could lead to a missed short-acting insulin dose reminder and a recommendation to correct a high glucose value. It does not impact insulin delivery, long-acting insulin dose reminders, or CGM alerts, and users can still use the pen itself to calculate a dose, deliver insulin, record the dose date/time, and view CGM data. This issue was identified during internal testing before release in the US but after release to OUS customers, no complaints or MDRs related to this recall have been reported. Use of the affected device may result hyperglycemia by failing to alert the user and delay treatment of diabetes.

Which lots are affected?

UDI-DI: 0763000B000122075 CFN/Software Version: MMT-8060/7.0.0, 7.1.0, 7.2.0 and 7.5.0 MMT-8061/7.0.0, 7.0.1, 7.1.0, 7.1.1, 7.2.0 and 7.5.0

Where was it sold?

International distribution to the countries of Argentina, Australia, Austria, Belgium, Bermuda, Chile, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Luxembourg, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.