CLASS II ONGOING Device recall · Reported 24 Dec 2025 · FDA Enforcement Report
InPen App recalled by Medtronic MiniMed
Medtronic MiniMed is recalling InPen App, distributed in 50 states. The recall was initiated on 13 Nov 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0958-2026
- FDA event ID
- 97904
- Recalling firm
- Medtronic MiniMed, Northridge, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 13 Nov 2025
- FDA classified
- 18 Dec 2025
- Reported
- 24 Dec 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Medical Software
Timeline
| 13 Nov 2025 | Recall initiated by company | |
| 18 Dec 2025 | Classified by FDA | +35 days |
| 24 Dec 2025 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
When app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't choose to allow notifications to override phone settings when on mute/Do Not Disturb, and previously set up override permission is deleted, so audible and vibratory notifications not received, which may lead to delayed insulin therapy, hyperglycemia.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
InPen App, Model/CFN Number: MMT-8061 (Android Users)
Where it was distributed
US: IL, AZ, MN, NE, KS, TN, CA, FL, TX, WA, MO, NC, NY, WI, IN, ME, CT, RI, MD, PA, OK, KY, MA, DE, UT, ID, VA, MI, OH, LA, NJ, IA, MT, HI, AR, OR, SC, AL, CO, WY, VT, NV, GA, MS, SD, AK, WV, NH, ND, NM. OUS: Austria, Belgium, Chile, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Ireland, Israel, Italy, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, South Africa, Spain, Sweden, Switzerland, United Kingdom
AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMaryland
Questions about this recall
Why was it recalled?
When app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't choose to allow notifications to override phone settings when on mute/Do Not Disturb, and previously set up override permission is deleted, so audible and vibratory notifications not received, which may lead to delayed insulin therapy, hyperglycemia.
Which lots are affected?
Software Versions: 7.5.0, 7.5.1, and 8.0.0. UDI-DI: 00763000974596
Where was it sold?
US: IL, AZ, MN, NE, KS, TN, CA, FL, TX, WA, MO, NC, NY, WI, IN, ME, CT, RI, MD, PA, OK, KY, MA, DE, UT, ID, VA, MI, OH, LA, NJ, IA, MT, HI, AR, OR, SC, AL, CO, WY, VT, NV, GA, MS, SD, AK, WV, NH, ND, NM. OUS: Austria, Belgium, Chile, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Ireland, Israel, Italy, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, South Africa, Spain, Sweden, Switzerland, United Kingdom
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Medtronic MiniMed
All 81| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | CareLink Clinic recalled by Medtronic MiniMed | Medtronic MiniMed | Device Malfunction | Nationwide |
| Device | CLASS II | The MiniMed Go app for iOS or MiniMed Go app recalled | Medtronic MiniMed | Device Malfunction | Nationwide |
| Device | CLASS II | MiniMed Flex Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Labeling Errors | Nationwide |
| Device | CLASS II | MiniMed Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Other | CA |
| Device | CLASS II | MiniMed Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Other | CA |
Other Medical Software recalls
All 935| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Metatarsophalangeal (MTP) Implant is a sterile recalled by restor3d | restor3d | Sterility | Nationwide |
| Device | CLASS II | Ion Torrent Genexus Dx recalled by Life Technologies Holdings Pte | Life Technologies Holdings Pte | Device Malfunction | Nationwide |
| Device | CLASS II | Mulitple-type osteoarticular pathogen nucleic acid IVD, kit recalled | BioFire Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS III | Kit: Urinary Clean Catch 20/cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Device Malfunction | CA |
| Device | CLASS II | Spectral CT and Spectral CT 7500 RT recalled by Philips Medical Systems | Philips Medical Systems | Device Malfunction | Nationwide |