CLASS II ONGOING Device recall · Reported 24 Jun 2026 · FDA Enforcement Report

Inspire Respiratory Sensing Lead recalled by Inspire Medical Systems

Inspire Medical Systems is recalling Inspire Respiratory Sensing Lead, distributed nationwide in the US. The recall was initiated on 11 May 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Code: Model No 4340 (USA only lead); UDI-DI [4340] 10855728005946; Inspire part number [4340] 900-014-003; serial numbers T18806, T45012, T45017, T49002, T67098, T67152, T69541, T69542, T69553, T69558, T69562 Model No 4340 (Global lead); UDI-DI [4340] 10810098650301; Inspire part number [4340] 900-014-005; serial numbers T91485, T91486, T91491, T91495, T91513, T91520, T91521, T91527, T91533, T91537, T91540, T91545, T96165, T96168, T96192, T96193, T96195, T96225, T96237, T98118, T98126, T98136, T98137, T98138, T98141, T98143, T98145, T98150, T98153, T98156, T98163, T98165, T98176, T98177, T98178, T98181, T98183, T98190, T96217
Quantity recalled50 impacted units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2515-2026
FDA event ID
98990
Recalling firm
Inspire Medical Systems, Golden Valley, MN
Classification
Class II
Status
Ongoing
Recall initiated
11 May 2026
FDA classified
18 Jun 2026
Reported
24 Jun 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

11 May 2026Recall initiated by company
18 Jun 2026Classified by FDA+38 days
24 Jun 2026Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire¿ Respiratory Sensing Lead (Model 4340) is designed to detect respiratory effort. The lead features a pressure sensitive membrane that converts the mechanical energy of respiration into an electrical signal. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG). Component: N/A

Where it was distributed

Worldwide distribution - US Nationwide and the country of Hong Kong.

Nationwide

Questions about this recall

Why was it recalled?

Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.

Which lots are affected?

Lot Code: Model No 4340 (USA only lead); UDI-DI [4340] 10855728005946; Inspire part number [4340] 900-014-003; serial numbers T18806, T45012, T45017, T49002, T67098, T67152, T69541, T69542, T69553, T69558, T69562 Model No 4340 (Global lead); UDI-DI [4340] 10810098650301; Inspire part number [4340] 900-014-005; serial numbers T91485, T91486, T91491, T91495, T91513, T91520, T91521, T91527, T91533, T91537, T91540, T91545, T96165, T96168, T96192, T96193, T96195, T96225, T96237, T98118, T98126, T98136, T98137, T98138, T98141, T98143, T98145, T98150, T98153, T98156, T98163, T98165, T98176, T98177, T98178, T98181, T98183, T98190, T96217

Where was it sold?

Worldwide distribution - US Nationwide and the country of Hong Kong.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 98990

This product is part of a recall event; the main page for the event is Inspire stimulation Lead recalled by Inspire Medical Systems.

ReportedClassRecallCompanyReasonWhere
CLASS IIInspire stimulation Lead recalled by Inspire Medical SystemsInspire Medical SystemsPotency & AssayNationwide

More recalls from Inspire Medical Systems

All 6

Other CPAP & Ventilators recalls

All 554