CLASS II ONGOING Device recall · Reported 27 May 2026 · FDA Enforcement Report

Integris-Allura system recalled by Philips Medical Systems

Philips Medical Systems Nederl is recalling Integris-Allura system, distributed nationwide in the US. The recall was initiated on 28 Apr 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes1. Integris CV Cesar-Powerpack-Visub-Nicol; Model: 722030; UDI-DI: N/A; Serial Numbers: 4888270, 59, 0, 33; 2. Integris Allura 15 & 12 (monoplane) (722043); UDI-DI: N/A; Serial Numbers: 365, 18151, 176, 239, 407, 426, 0; 3. INTEGRIS Allura 15-12 (biplane) (722044); UDI-DI: N/A; Serial Numbers: 86322, 30, 66; 4. Integris-Allura 9 (722018); UDI-DI: N/A; Serial Numbers: 26, 85; 5. Integris-Allura 9 (Biplane) (722021); UDI-DI: N/A; Serial numbers: 22, 75, 25, 108; 6. Integris H5000 C / Allura 9C (722016); UDI-DI: N/A; Serial Numbers: 4059001008, 4807240, S01H003553/000001;
Quantity recalled25 units (11 US, 14 OUS)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2229-2026
FDA event ID
98801
Recalling firm
Philips Medical Systems Nederl, Best
Classification
Class II
Status
Ongoing
Recall initiated
28 Apr 2026
FDA classified
21 May 2026
Reported
27 May 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

28 Apr 2026Recall initiated by company
21 May 2026Classified by FDA+23 days
27 May 2026Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Integris-Allura system; System Code Description (Model Numbers): Integris CV Cesar-Powerpack-Visub-Nicol (722030), Integris Allura 15 & 12 (monoplane) (722043), INTEGRIS Allura 15-12 (biplane) (722044), Integris-Allura 9 (722018), Integris-Allura 9 (Biplane) (722021), Integris CV Cesar-Powerpack (722030), Integris H5000 C / Allura 9C (722016);

Where it was distributed

Domestic: Nationwide; International: Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Martinique, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam, Yemen;

Nationwide Georgia

Questions about this recall

Why was it recalled?

The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.

Which lots are affected?

1. Integris CV Cesar-Powerpack-Visub-Nicol; Model: 722030; UDI-DI: N/A; Serial Numbers: 4888270, 59, 0, 33; 2. Integris Allura 15 & 12 (monoplane) (722043); UDI-DI: N/A; Serial Numbers: 365, 18151, 176, 239, 407, 426, 0; 3. INTEGRIS Allura 15-12 (biplane) (722044); UDI-DI: N/A; Serial Numbers: 86322, 30, 66; 4. Integris-Allura 9 (722018); UDI-DI: N/A; Serial Numbers: 26, 85; 5. Integris-Allura 9 (Biplane) (722021); UDI-DI: N/A; Serial numbers: 22, 75, 25, 108; 6. Integris H5000 C / Allura 9C (722016); UDI-DI: N/A; Serial Numbers: 4059001008, 4807240, S01H003553/000001;

Where was it sold?

Domestic: Nationwide; International: Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Martinique, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam, Yemen;

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 98801

This product is part of a recall event; the main page for the event is Azurion system recalled by Philips Medical Systems.

ReportedClassRecallCompanyReasonWhere
CLASS IIAzurion system recalled by Philips Medical SystemsPhilips Medical Systems NederlPackaging DefectsNationwide

More recalls from Philips Medical Systems Nederl

All 200

Other Other Medical Devices recalls

All 13,950