CLASS II ONGOING Device recall · Reported 10 Dec 2025 · FDA Enforcement Report

IntelliVue Patient Monitor MX600 recalled by Philips North America

Philips North America is recalling IntelliVue Patient Monitor MX600, distributed nationwide in the US. The recall was initiated on 31 Oct 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesProduct Number: 865242; UDI-DI: 00884838029095; All Serial No.
Quantity recalled1,913,441 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0866-2026
FDA event ID
97917
Recalling firm
Philips North America, Cambridge, MA
Classification
Class II
Status
Ongoing
Recall initiated
31 Oct 2025
FDA classified
2 Dec 2025
Reported
10 Dec 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Patient Monitors

Timeline

31 Oct 2025Recall initiated by company
2 Dec 2025Classified by FDA+32 days
10 Dec 2025Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Potential issue where the IntelliVue monitors did not alarm.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

IntelliVue Patient Monitor MX600. Product Number: 865242.

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia, Botswana, Brazil, Brunei, Bulgaria, Burundi, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Equat Guinea, Estonia, Ethiopia, Faroe Islands, Finland, French Polynesia, France, French Guiana, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guadeloupe, Guam, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea Rep, Kuwait, Lao, Latvia, Lebanon, Lesotho, Liberia, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, Netherlands Ant, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, , R¿union, Romania, Russia, Saint Vincent, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, St. Pierre Mqln, Suriname, Sweden, Switzerland, Syrian Arab Rep, Taiwan, Tajikistan, Thailand, Trinidad Tobago, Tunisia, T¿rkiye, Turkmenistan, Uganda, Ukraine, United Kingdom, United States, Uruguay, Utd Arab Emir, Uzbekistan, Venezuela, Viet Nam, Virgin Isl (Br), Virgin Isl (US), White Russia, Yemen, Zambia, Zimbabwe.

Nationwide GeorgiaPuerto Rico

Questions about this recall

Why was it recalled?

Potential issue where the IntelliVue monitors did not alarm.

Which lots are affected?

Product Number: 865242; UDI-DI: 00884838029095; All Serial No.

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia, Botswana, Brazil, Brunei, Bulgaria, Burundi, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Equat Guinea, Estonia, Ethiopia, Faroe Islands, Finland, French Polynesia, France, French Guiana, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guadeloupe, Guam, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea Rep, Kuwait, Lao, Latvia, Lebanon, Lesotho, Liberia, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, Netherlands Ant, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, , R¿union, Romania, Russia, Saint Vincent, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, St. Pierre Mqln, Suriname, Sweden, Switzerland, Syrian Arab Rep, Taiwan, Tajikistan, Thailand, Trinidad Tobago, Tunisia, T¿rkiye, Turkmenistan, Uganda, Ukraine, United Kingdom, United States, Uruguay, Utd Arab Emir, Uzbekistan, Venezuela, Viet Nam, Virgin Isl (Br), Virgin Isl (US), White Russia, Yemen, Zambia, Zimbabwe.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

21 other products · FDA event 97917

This product is part of a recall event; the main page for the event is IntelliVue MP90 recalled by Philips North America.

ReportedClassRecallCompanyReasonWhere
CLASS IIIntelliVue MP90 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue MP30 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue Patient Monitor MX800 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue Patient Monitor MX700 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue Patient Monitor MX500 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue Patient Monitor MX850 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue MP2 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue MP50 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue MP80 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue Patient Monitor MX100 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue MP40 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue Multi Measurement Server X2 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue MP60 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue MP5 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue MP20 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue Patient Monitor MX400 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue Multi-Measurement Module X3 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue Patient Monitor MX550 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue Patient Monitor MX750 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue Patient Monitor MX450 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide
CLASS IIIntelliVue MP70 recalled by Philips North AmericaPhilips North AmericaDevice MalfunctionNationwide

More recalls from Philips North America

All 531

Other Patient Monitors recalls

All 1,013