CLASS II Device recall · Reported 23 Mar 2016 · FDA Enforcement Report
Intra operative MRI Systems: Neuro Ii-se , Neuro Iii-sv recalled
Deerfield Imaging is recalling Intra operative MRI Systems: Neuro Ii-se , Neuro Iii-sv, iMRI 1.5T A, and iMRI 3T S, distributed in 19 states. The recall was initiated on 10 Feb 2016 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1098-2016
- FDA event ID
- 73421
- Recalling firm
- Deerfield Imaging, Minnetonka, MN
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 10 Feb 2016
- FDA classified
- 11 Mar 2016
- Reported
- 23 Mar 2016
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Imaging Equipment
Timeline
| 10 Feb 2016 | Recall initiated by company | |
| 11 Mar 2016 | Classified by FDA | +30 days |
| 23 Mar 2016 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Emergency helium venting lines (i.e. quench lines) may not have been properly installed.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Intra operative MRI Systems: Neuro II-SE , Neuro III-SV, iMRI 1.5T A, and iMRI 3T S. Indicated for use as a magnetic resonance diagnostic device (MRDD).
Where it was distributed
Worldwide Distribution -- US, including the states of CA, CT, FL, GA, IL, IN, MD, MA, MN, MO, NH, NC, OH, OR, PA, TX, UT, VA, and WI; and ,countries of CHINA, CANADA, AUSTRALIA, FRANCE, JAPAN, and GERMANY.
CaliforniaConnecticutFloridaGeorgiaIllinoisIndianaMassachusettsMarylandMinnesotaMissouriNorth CarolinaNew HampshireOhioOregonPennsylvaniaTexasUtahVirginiaWisconsin
Questions about this recall
Why was it recalled?
Emergency helium venting lines (i.e. quench lines) may not have been properly installed.
Which lots are affected?
70036, 10076-21, M00004, 10073-01, 23529, 40481, M00152, M00005, 30445, M00169, 060014, 10043-00, 30823, 10004680, 10050-01, 40662, 30935, 31084, 31203, 10002221, 40510, 40613, 30719, 31046, 30910, 30832, 30623, 31213, 10004152, 10072-01, 31250, 40815, 31049, 30068, 40412, M0038, M00106, M00367, 10035-01, 40475, 40495, 10007498, 10005946, 40263, 40688, 10058-00, 40619, 31085, and 10005639.
Where was it sold?
Worldwide Distribution -- US, including the states of CA, CT, FL, GA, IL, IN, MD, MA, MN, MO, NH, NC, OH, OR, PA, TX, UT, VA, and WI; and ,countries of CHINA, CANADA, AUSTRALIA, FRANCE, JAPAN, and GERMANY.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Deerfield Imaging
All 21| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Trumpf Mount Clamp Assembly (table adapter component) recalled | Deerfield Imaging | Other | Nationwide |
| Device | CLASS II | Operating room table recalled by Deerfield Imaging | Deerfield Imaging | Device Malfunction | Nationwide |
| Device | CLASS II | Operating room table recalled by Deerfield Imaging | Deerfield Imaging | Device Malfunction | Nationwide |
| Device | CLASS II | Operating room table recalled by Deerfield Imaging | Deerfield Imaging | Device Malfunction | Nationwide |
| Device | CLASS II | iMRX MRI and Angiography System recalled by Deerfield Imaging | Deerfield Imaging | Device Malfunction | CT, MA, VA |
Other Imaging Equipment recalls
All 2,347| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |