CLASS II ONGOING Device recall · Reported 7 Oct 2026 · FDA Enforcement Report

Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065 recalled

Percussionaire is recalling Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065, distributed nationwide in the US. The recall was initiated on 18 Aug 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesProduct Number/UDI-DI: S00065/00849436000679, 00849436000709; P5-UC/00849436000686 In-Line Valve Instruction Manual: P20174 Rev D
Quantity recalled61,858

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-3319-2026
FDA event ID
99617
Recalling firm
Percussionaire, Sandpoint, ID
Classification
Class II
Status
Ongoing
Recall initiated
18 Aug 2026
FDA classified
28 Sep 2026
Reported
7 Oct 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

18 Aug 2026Recall initiated by company
28 Sep 2026Classified by FDA+41 days
7 Oct 2026Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

Product as listed by the FDA

Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065; Phasitron 5 Breathing Circuit with Unified Connector, Product Number: P5-UC, used with In-Line Valve Instruction Manual

Where it was distributed

Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasLouisianaMassachusettsMaryland

Questions about this recall

Why was it recalled?

Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

Which lots are affected?

Product Number/UDI-DI: S00065/00849436000679, 00849436000709; P5-UC/00849436000686 In-Line Valve Instruction Manual: P20174 Rev D

Where was it sold?

Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 13 Oct 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

6 other products · FDA event 99617

This product is part of a recall event; the main page for the event is Intrapulmonary Percussive Ventilation, IPV HC recalled by Percussionaire.

More recalls from Percussionaire

All 15

Other Other Medical Devices recalls

All 13,884