CLASS II Device recall · Reported 24 Feb 2021 · FDA Enforcement Report

IOLMaster recalled by Carl Zeiss Meditec AG

Carl Zeiss Meditec is recalling IOLMaster 700, distributed in 39 states. The recall was initiated on 7 Dec 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel: IOLMaster 700 Catalog Number device: 000000-1932-169
Quantity recalled613 devices

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1133-2021
FDA event ID
87055
Recalling firm
Carl Zeiss Meditec, Jena, N/A
Classification
Class II
Status
Terminated
Recall initiated
7 Dec 2020
FDA classified
18 Feb 2021
Reported
24 Feb 2021
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

7 Dec 2020Recall initiated by company
18 Feb 2021Classified by FDA+73 days
24 Feb 2021Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

When using software 1.90.2.09 or 1.90.8.06 and using modality worklist functionality for patient data transfer, the selection of the patient in the patient list may not match the patient information displayed on the right side of the screen

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

IOLMaster 700

Where it was distributed

U.S.: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: Austria, Australia, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, France, United Kingdom, Guatemala, Hong Kong, Indonesia, Ireland, India, Italy, Japan, Kuwait, Malaysia, Netherlands, New Zealand, Philippines, Poland, Portugal, Russia, Sweden, Singapore, Slovakia, Thailand, Taiwan, and South Africa.

AlabamaArkansasArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaIowaIllinoisIndianaKansasLouisianaMassachusettsMarylandMaineMichiganMissouriMississippi

Questions about this recall

Why was it recalled?

When using software 1.90.2.09 or 1.90.8.06 and using modality worklist functionality for patient data transfer, the selection of the patient in the patient list may not match the patient information displayed on the right side of the screen

Which lots are affected?

Model: IOLMaster 700 Catalog Number device: 000000-1932-169

Where was it sold?

U.S.: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: Austria, Australia, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, France, United Kingdom, Guatemala, Hong Kong, Indonesia, Ireland, India, Italy, Japan, Kuwait, Malaysia, Netherlands, New Zealand, Philippines, Poland, Portugal, Russia, Sweden, Singapore, Slovakia, Thailand, Taiwan, and South Africa.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Carl Zeiss Meditec

All 21

Other Other Medical Devices recalls