CLASS II ONGOING Device recall · Reported 10 Jul 2024 · FDA Enforcement Report

IsoLoc Prostate Immobilization Treatment Balloon Device Kit recalled

Angiodynamics is recalling IsoLoc Prostate Immobilization Treatment Balloon Device Kit, distributed nationwide in the US. The recall was initiated on 21 May 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUPN/Catalog: ISO-100; UDI/DI: 10851546007100 (Box) and 00851546007103 (Pouch); Lot number: 5402.
Quantity recalled1,500 units (75 kits with 20 units per kit)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2267-2024
FDA event ID
94773
Recalling firm
Angiodynamics, Queensbury, NY
Classification
Class II
Status
Ongoing
Recall initiated
21 May 2024
FDA classified
3 Jul 2024
Reported
10 Jul 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Labeling Errors

Timeline

21 May 2024Recall initiated by company
3 Jul 2024Classified by FDA+43 days
10 Jul 2024Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A single batch of ISO-100 IsoLoc Gas-Release Endorectal Balloons is mislabeled. The affected ISO-100 balloon catheter has a nominal fill volume of 100cc/ml but the inner pouch identifies the product as an ISO-60 (fill volume of 60cc/ml).

Product as listed by the FDA

IsoLoc Prostate Immobilization Treatment Balloon Device Kit, Catalog/UPN: ISO-100

Where it was distributed

US Nationwide distribution in the states of AZ, FL, IL, MO, MS, NJ, NM, NY, PA, TX, WI.

Nationwide ArizonaFloridaIllinoisMissouriMississippiNew JerseyNew MexicoNew YorkPennsylvaniaTexasWisconsin

Questions about this recall

Why was it recalled?

A single batch of ISO-100 IsoLoc Gas-Release Endorectal Balloons is mislabeled. The affected ISO-100 balloon catheter has a nominal fill volume of 100cc/ml but the inner pouch identifies the product as an ISO-60 (fill volume of 60cc/ml).

Which lots are affected?

UPN/Catalog: ISO-100; UDI/DI: 10851546007100 (Box) and 00851546007103 (Pouch); Lot number: 5402.

Where was it sold?

US Nationwide distribution in the states of AZ, FL, IL, MO, MS, NJ, NM, NY, PA, TX, WI.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.